IMMUCOR/GAMMA CSW 100 6510/8441-10 *

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed with the FDA on 2007-03-25 for IMMUCOR/GAMMA CSW 100 6510/8441-10 * manufactured by Tricontinent.

Event Text Entries

[592449] 1. Biohazardous waste was aerosolized from the waste container. 2. Unit puts waste into the waste container with the waste sensor unplugged. The immucor/gamma rep was on site to train laboratory personnel on the capture system. She was demonstrating the waste bottle sensor features. The waste bottle was almost full -as described in section 2. 6 page 6 of the operation and maintenance manual-. A full tray of wells was loaded to wash. Start was pressed. The system began washing the strips. After washing a number of strips, biohazardous waste began spewing out the top of the waste bottle. The system continued to wash. User had to press the "stop" button before the system would stop aerosolizing biohazardous waste from the top of the waste bottle. With the waste sensor unplugged or malfunctioning -waste bottle level is not known- the system will prime and rinse the probes causing waste to go into a waste bottle that may already be full.
Patient Sequence No: 1, Text Type: D, B5


MAUDE Entry Details

Report NumberMW1042345
MDR Report Key832733
Date Received2007-03-25
Date of Report2007-03-25
Date of Event2007-02-23
Date Added to Maude2007-04-02
Event Key0
Report Source CodeVoluntary report
Manufacturer LinkN
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Reporter OccupationOTHER HEALTH CARE PROFESSIONAL
Health Professional3
Initial Report to FDA3
Report to FDA3
Event Location3
Single Use0
Previous Use Code0
Event Type3
Type of Report3

Device Sequence Number: 1

Brand NameIMMUCOR/GAMMA CSW 100
Generic NameCAPTURE STRIP WELL WASHER
Product CodeFLH
Date Received2007-03-25
Model Number6510/8441-10
Catalog Number*
Lot Number*
ID Number*
OperatorHEALTH PROFESSIONAL
Device AvailabilityY
Device Age*
Implant FlagN
Date Removed*
Device Sequence No1
Device Event Key820052
ManufacturerTRICONTINENT
Manufacturer Address12555 LOMA RICA DR GRASS VALLEY CA 95945 US

Device Sequence Number: 2

Brand NameGAMMA CSW 100
Generic NameCAPTURE STRIP WELL WASHER
Product CodeFLH
Date Received2007-03-25
Model Number8441-10
Catalog Number*
Lot Number*
ID Number*
OperatorHEALTH PROFESSIONAL
Device AvailabilityY
Device Age*
Implant FlagN
Date Removed*
Device Sequence No2
Device Event Key820163
ManufacturerTRICONTINENT
Manufacturer Address12555 LOMA RICA DR GRASS VALLEY CA 95945 US


Patients

Patient NumberTreatmentOutcomeDate
10 2007-03-25

© 2026 FDA.report