VOLDYNE 5000 VOLUMETRIC EXERCISER 8884719009

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed from a company representative,health report with the FDA on 2019-02-12 for VOLDYNE 5000 VOLUMETRIC EXERCISER 8884719009 manufactured by Teleflex Medical.

Event Text Entries

[136048132] Qn# (b)(4). The customer returned one unit 8884719009 voldyne 5000 volumetric exerciser for investigation. The returned breathing exerciser was visually examined and no defects or anomalies were observed. The returned volumetric exerciser was able to pass a functional inspection when connected to a vacuumed air. The piston and yellow flow cup were both able to move freely within the device. A device history record review was performed on the received sample lot number 73j1800456, with no evidence of a manufacturing related issue found. The reported complaint of "yellow float sticks during use" was not confirmed based upon the sample received. There were no functional issues found with the returned sample. A corrective action is not required at this time as there were no functional issues found with the returned sample. (continued). Other remarks: the ifu for this product, l06121, was reviewed as a part of this complaint investigation. The ifu instructs the end user, "exhale normally. Then place lips tightly around mouthpiece. Inhale slowly to raise the white piston in the chamber. Inhale slowly to raise the piston in the chamber. When inhaling, maintain top of the yellow flow cup in the 'best' flow range. " "continue inhaling and try to raise piston to prescribed level. " telefelex will continue to monitor customer feedback for complaints of this nature.
Patient Sequence No: 1, Text Type: N, H10


[136048133] Customer complaint states: "rt reported that the yellow float on the left side is sticking. " no patient harm reported. Patient condition reported as fine.
Patient Sequence No: 1, Text Type: D, B5


MAUDE Entry Details

Report Number3003898360-2019-00079
MDR Report Key8330827
Report SourceCOMPANY REPRESENTATIVE,HEALTH
Date Received2019-02-12
Date of Report2019-01-16
Date of Event2019-01-16
Date Mfgr Received2019-01-16
Date Added to Maude2019-02-12
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Reporter OccupationOTHER HEALTH CARE PROFESSIONAL
Health Professional3
Initial Report to FDA3
Report to FDA3
Event Location3
Manufacturer ContactLINDA WOODALL
Manufacturer Street3015 CARRINGTON MILL BLVD
Manufacturer CityMORRISVILLE 27560
Manufacturer CountryUS
Manufacturer Postal27560
Manufacturer Phone9196942566
Manufacturer G1TELEFLEX MEDICAL
Manufacturer StreetRANCHO EL DESCANSO
Manufacturer CityTECATE 21478
Manufacturer CountryMX
Manufacturer Postal Code21478
Single Use3
Previous Use Code3
Event Type3
Type of Report3

Device Details

Brand NameVOLDYNE 5000 VOLUMETRIC EXERCISER
Generic NameSPIROMETER, THERAPEUTIC (INCEN
Product CodeBWF
Date Received2019-02-12
Returned To Mfg2019-01-23
Catalog Number8884719009
Lot NumberUNKNOWN
Device AvailabilityR
Device Eval'ed by MfgrY
Device Sequence No1
Device Event Key0
ManufacturerTELEFLEX MEDICAL
Manufacturer AddressRESEARCH TRIANGLE PARK NC


Patients

Patient NumberTreatmentOutcomeDate
10 2019-02-12