VISCOUS FLUID CONTROL TUBING SET (SILICONE OIL INJ/EXT) 8065750957

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed from a foreign,health professional report with the FDA on 2019-02-12 for VISCOUS FLUID CONTROL TUBING SET (SILICONE OIL INJ/EXT) 8065750957 manufactured by Alcon Research, Ltd. - Houston.

Event Text Entries

[135876728] Investigation, including root cause analysis, is in progress. A supplemental mdr will be filed as necessary in accordance with 21 cfr 803. 56 when additional reportable information becomes available. (b)(4).
Patient Sequence No: 1, Text Type: N, H10


[135876729] A physician reported that the console displayed "too much pressure in the syringe for silicone oil" during a cataract/vitrectomy procedure. The pressure was reported to be about 70-75mm hg (millimeters of mercury). The cannula popped off the syringe. The product was replaced and the procedure was completed. The reporter informed that the patient experienced bleeding because of the change of the pressure. The patients current condition is unknown.
Patient Sequence No: 1, Text Type: D, B5


MAUDE Entry Details

Report Number1644019-2019-00018
MDR Report Key8331234
Report SourceFOREIGN,HEALTH PROFESSIONAL
Date Received2019-02-12
Date of Report2019-05-29
Date of Event2019-02-11
Date Mfgr Received2019-05-10
Date Added to Maude2019-02-12
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Health Professional3
Initial Report to FDA3
Report to FDA3
Event Location3
Manufacturer ContactMS. CINDY MILAM
Manufacturer Street6201 SOUTH FREEWAY MAIL STOP AB2-6
Manufacturer CityFORT WORTH TX 76134
Manufacturer CountryUS
Manufacturer Postal76134
Manufacturer Phone8176152231
Manufacturer G1ALCON RESEARCH, LLC - HOUSTON
Manufacturer Street9965 BUFFALO SPEEDWAY
Manufacturer CityHOUSTON TX 77054
Manufacturer CountryUS
Manufacturer Postal Code77054
Single Use3
Previous Use Code3
Removal Correction NumberNA
Event Type3
Type of Report3

Device Details

Brand NameVISCOUS FLUID CONTROL TUBING SET (SILICONE OIL INJ/EXT)
Generic NameTUBING, NONINVASIVE
Product CodeGAZ
Date Received2019-02-12
Model NumberNA
Catalog Number8065750957
Lot NumberASKU
OperatorHEALTH PROFESSIONAL
Device AvailabilityN
Device AgeDA
Device Eval'ed by MfgrR
Device Sequence No1
Device Event Key0
ManufacturerALCON RESEARCH, LTD. - HOUSTON
Manufacturer Address9965 BUFFALO SPEEDWAY HOUSTON TX 77054 US 77054


Patients

Patient NumberTreatmentOutcomeDate
101. Other 2019-02-12

© 2026 FDA.report
This site is not affiliated with or endorsed by the FDA.