PRONE HEADREST 40061900

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed from a foreign,health professional,u report with the FDA on 2019-02-12 for PRONE HEADREST 40061900 manufactured by Holger Ullrich.

Event Text Entries

[135912385] (b)(4). Pressure ulcers can be caused due to different influences (preexisting medical conditions, patient age, duration of the position, unfavorable positioning of the patient, etc. ). No damage or malfunction of the used gel pad was reported. In the instructions for use (ifu) the user is warned regarding the risks related to improper patient positioning. Quote ifu: "warning! Risk of injury! Improper patient positioning may cause health damage (e. G. Decubitus). Position the patient correctly and keep under constant observation. " no malfunction of the (b)(4) - maquet device was reported. We assume that in this case several unfavorable factors came together and thus led to the described injury. Since an injury was reported to (b)(4) - maquet we decided to report tis case to the competent authorities. (b)(4) -maquet provides product failure investigation, analysis and resolution for the device described in this report.
Patient Sequence No: 1, Text Type: N, H10


[135912386] It was reported that a patient developed decubitus on the chin while a (b)(4) - maquet gel pad was in use. The patient was under anaesthesia and in a prone position. Manufacturer reference # (b)(4).
Patient Sequence No: 1, Text Type: D, B5


MAUDE Entry Details

Report Number8010652-2019-00005
MDR Report Key8331826
Report SourceFOREIGN,HEALTH PROFESSIONAL,U
Date Received2019-02-12
Date of Report2019-09-25
Date of Event2019-02-07
Date Mfgr Received2019-08-30
Date Added to Maude2019-02-12
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Health Professional3
Initial Report to FDA3
Report to FDA3
Event Location3
Manufacturer StreetMAQUET GMBH KEHLER STRASSE 31,
Manufacturer CityRASTATT 76437
Manufacturer CountryGM
Manufacturer Postal76437
Manufacturer G1HOLGER ULLRICH
Manufacturer StreetMAQUET GMBH KEHLER STRASSE 31,
Manufacturer CityRASTATT 76437
Manufacturer CountryGM
Manufacturer Postal Code76437
Single Use3
Previous Use Code3
Event Type3
Type of Report3

Device Details

Brand NamePRONE HEADREST
Generic NameSURGICAL TABLE CUSHION
Product CodeLWG
Date Received2019-02-12
Model Number40061900
Catalog Number40061900
OperatorHEALTH PROFESSIONAL
Device AvailabilityY
Device AgeDA
Device Eval'ed by MfgrY
Device Sequence No1
Device Event Key0
ManufacturerHOLGER ULLRICH
Manufacturer AddressMAQUET GMBH KEHLER STRASSE 31, RASTATT 76437 GM 76437


Patients

Patient NumberTreatmentOutcomeDate
101. Other 2019-02-12

© 2026 FDA.report
This site is not affiliated with or endorsed by the FDA.