OCEAN VIP W/TOILET PAN HOLDER AND PAN SOFT SEAT 9153650627 NA:1470793KIT

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed from a company representative report with the FDA on 2019-02-13 for OCEAN VIP W/TOILET PAN HOLDER AND PAN SOFT SEAT 9153650627 NA:1470793KIT manufactured by Aquatec Operations Gmbh.

Event Text Entries

[136036110] It was reported to invacare that a resident of a nursing facility was taken into the washroom by her daughter in a wheelchair. Staff transferred the patient into the shower commode in the washroom. After toileting, the daughter was wheeling the mother in the shower commode. As the shower chair crossed over the threshold, the wheel of the shower commode caught on the threshold. The mother pitched forward and fell out the chair face forward, hitting the floor. She was taken to the hospital. The resident sustained an inter-cranial hemorrhage (not major) and was monitored at the hospital for 24 hours, then released without admission with no further concern. The invacare territory business manager visited the facility where it was confirmed that there was no malfunction with the shower commode chair. The resident was wheeled out of the washroom in the shower commode chair, not her wheelchair. The user manual states that this product is to be used only on flat smooth surfaces. The resident's daughter confirmed that her mother was not positioned properly. The mother is a large heavy woman and was not sitting all the way back in the chair, thus the center of gravity was out of balance.
Patient Sequence No: 1, Text Type: N, H10


[136036111] An invacare territory business manager (tbm) reported that a resident was being wheeled out of a washroom at a facility by a family member in an ocean vip w/toilet pan holder and pan soft seat (tilt-in-space shower commode chair) when the caster wheel hit the threshold causing the resident to fall from the chair.
Patient Sequence No: 1, Text Type: D, B5


MAUDE Entry Details

Report Number3007231105-2019-00002
MDR Report Key8335280
Report SourceCOMPANY REPRESENTATIVE
Date Received2019-02-13
Date of Report2019-01-22
Date of Event2019-01-19
Date Mfgr Received2019-01-22
Device Manufacturer Date2018-10-01
Date Added to Maude2019-02-13
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Health Professional3
Initial Report to FDA3
Report to FDA3
Event Location3
Manufacturer ContactMR. JASON FIEST
Manufacturer StreetONE INVACARE WAY
Manufacturer CityELYRIA OH 44035
Manufacturer CountryUS
Manufacturer Postal44035
Manufacturer Phone8003336900
Manufacturer G1AQUATEC OPERATIONS GMBH
Manufacturer StreetALEMANNENSTRASSE 10
Manufacturer CityISNY 88316
Manufacturer CountryGM
Manufacturer Postal Code88316
Single Use3
Previous Use Code3
Event Type3
Type of Report3

Device Details

Brand NameOCEAN VIP W/TOILET PAN HOLDER AND PAN SOFT SEAT 9153650627
Generic NameDAILY ACTIVITY DEVICE
Product CodeILS
Date Received2019-02-13
Model NumberNA:1470793KIT
OperatorLAY USER/PATIENT
Device AvailabilityN
Device Eval'ed by MfgrR
Device Sequence No1
Device Event Key0
ManufacturerAQUATEC OPERATIONS GMBH
Manufacturer AddressALEMANNENSTRASSE 10 ISNY 88316 GM 88316


Patients

Patient NumberTreatmentOutcomeDate
101. Other 2019-02-13

© 2026 FDA.report