ROSCH-UCHIDA TRANSJUGULAR LIVER ACCESS SET N/A RUPS-100

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed from a company representative,foreig report with the FDA on 2019-02-14 for ROSCH-UCHIDA TRANSJUGULAR LIVER ACCESS SET N/A RUPS-100 manufactured by Cook Inc.

Event Text Entries

[136295457] Concomitant medical products: terumo sheath, johnson & johnson balloon, gore viatorr stent. Occupation: agent. Pma/510(k) #: exempt. This report includes information known at this time. A follow up report will be submitted should additional relevant information become available.
Patient Sequence No: 1, Text Type: N, H10


[136295458] It was reported an rosch-uchida transjugular liver access set was used in a male patient with cirrhosis of the liver and portal hypertension. As reported, during the advancement of the sheath, the proximal end was detached from the check-flo assembly valve. The operator removed the damaged sheath and was able to complete the procedure successfully with a new rosch-uchida transjugular liver access set. The complaint device was archived at the manufacturer to aid in the investigation. Excessive biomatter was present all over the device and the check-flo assembly was confirmed to be detached from the sheath. Other than the check-flo assembly detachment, no other damage was noted to the device. The inner coiling of the device was noted to be unraveled at the proximal end where the separation occurred. A portion of the sheath tubing was found within the hub. As reported, the patient did not experience any adverse effects or require any additional procedures due to this occurrence.
Patient Sequence No: 1, Text Type: D, B5


MAUDE Entry Details

Report Number1820334-2019-00180
MDR Report Key8338706
Report SourceCOMPANY REPRESENTATIVE,FOREIG
Date Received2019-02-14
Date of Report2019-03-05
Date of Event2019-01-30
Date Mfgr Received2019-03-04
Device Manufacturer Date2018-11-27
Date Added to Maude2019-02-14
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Health Professional3
Initial Report to FDA3
Report to FDA3
Event Location3
Manufacturer ContactMR. LARRY POOL
Manufacturer Street750 DANIELS WAY
Manufacturer CityBLOOMINGTON IN 47404
Manufacturer CountryUS
Manufacturer Postal47404
Manufacturer Phone8123392235
Manufacturer G1COOK INC
Manufacturer Street750 DANIELS WAY
Manufacturer CityBLOOMINGTON IN 47404
Manufacturer CountryUS
Manufacturer Postal Code47404
Single Use3
Previous Use Code3
Event Type3
Type of Report3

Device Details

Brand NameROSCH-UCHIDA TRANSJUGULAR LIVER ACCESS SET
Generic NameGBZ CATHETER, CHOLANGIOGRAPHY
Product CodeGBZ
Date Received2019-02-14
Returned To Mfg2019-02-07
Model NumberN/A
Catalog NumberRUPS-100
Lot Number9335751
OperatorHEALTH PROFESSIONAL
Device AvailabilityR
Device AgeDA
Device Eval'ed by MfgrY
Device Sequence No1
Device Event Key0
ManufacturerCOOK INC
Manufacturer Address750 DANIELS WAY BLOOMINGTON IN 47404 US 47404


Patients

Patient NumberTreatmentOutcomeDate
10 2019-02-14

© 2026 FDA.report
This site is not affiliated with or endorsed by the FDA.