PRISMAFLO IIS PF2-WP33

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed with the FDA on 2019-02-15 for PRISMAFLO IIS PF2-WP33 manufactured by Stihler Electronic Gmbh.

Event Text Entries

[136515443] Based on the investigation is that the heating wire of the heating profile was internally damaged due to external mechanical stress additionally the heating profile was partly covered by patients arm and flank. There is also a strong doubt if the cleaning and disinfecting methods (only with alcohol-based disinfectant agents or ready-to-use disinfectants) for the heating profile were followed correctly (according the instructions for use). The combination of these factors must have lead to the local overheating of the heating profile and consequently to the two burn marks on right arm and flank. The burn marks were small and not judged a serious injury. Because of the characteristics of this event, the manufacturer judges that this event fits in the recall of this product that addresses the potential failure. As soon as the healthcare provider responds to the "urgent medical device correction" form from stihler electronic the affected heating profiles will be replaced.
Patient Sequence No: 1, Text Type: N, H10


[136515444] It was reported that during continuous renal replacement therapy (crrt) using a blood warmer prismaflo iis, the nurse noticed that the heating profile of the blood warmer was hot and smoking. The patient suffered at least two second degree burns. Treatment was discontinued.
Patient Sequence No: 1, Text Type: D, B5


MAUDE Entry Details

Report Number9617473-2017-00001
MDR Report Key8342338
Date Received2019-02-15
Date Mfgr Received2017-03-01
Date Added to Maude2019-02-15
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag0
Product Problem Flag3
Reprocessed and Reused Flag3
Health Professional0
Initial Report to FDA0
Report to FDA3
Event Location3
Manufacturer ContactMR. MICHAEL SCHELLING
Manufacturer StreetJULIUS-HOELDER-STRASSE 36
Manufacturer CitySTUTTGART, 70597
Manufacturer CountryGM
Manufacturer Postal70597
Manufacturer G1STIHLER ELECTRONIC GMBH
Manufacturer StreetJULIUS-HOELDER-STRASSE 36
Manufacturer CitySTUTTGART, 70597
Manufacturer CountryGM
Manufacturer Postal Code70597
Single Use3
Remedial ActionRC
Previous Use Code3
Removal Correction NumberZ-0759-2019
Event Type3
Type of Report0

Device Details

Brand NamePRISMAFLO IIS
Generic NameBLOOD WARMER
Product CodeKOC
Date Received2019-02-15
Model NumberPF2-WP33
Catalog NumberPF2-WP33
Device AvailabilityN
Device Eval'ed by MfgrR
Device Sequence No1
Device Event Key0
ManufacturerSTIHLER ELECTRONIC GMBH
Manufacturer AddressJULIUS-HOELDER-STRASSE 36 STUTTGART, 70597 GM 70597


Patients

Patient NumberTreatmentOutcomeDate
10 2019-02-15

© 2026 FDA.report
This site is not affiliated with or endorsed by the FDA.