PRISMAFLO II OR PRISMAFLO IIS NO INFORMATION

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed with the FDA on 2019-02-15 for PRISMAFLO II OR PRISMAFLO IIS NO INFORMATION manufactured by Stihler Electronic Gmbh.

Event Text Entries

[136520766] Further investigations are not possible as the defective heating profile was not available for investigation and s/n of defective heating profile was not transmitted. Should additional relevant information become available, a supplement report will be submitted. Because of the characteristics of this event the manufacturer judges that this event fits in the recall of this product that addresses the potential failure. The indication that pca was "sapped" is not possible as the applied part of product is classified "cf". As soon as the healthcare provider responds to the "urgent medical device correction" form from stihler electronic the affected heating profile will be replaced. There was no patient injury or medical intervention associated with this event.
Patient Sequence No: 1, Text Type: N, H10


[136520767] It was reported that during continous renal replacement therapy (crrt) using a blood warmer (prsimaflo ii or prismaflo iis). The user noticed a small bubble on the surface of the heating profile. The bubble was glowing red and issuing a small amount of smoke. Treatment was discontinued. There was no report of patient injury or medical intervention associated with this event. No additional information is available.
Patient Sequence No: 1, Text Type: D, B5


MAUDE Entry Details

Report Number9617473-2018-00003
MDR Report Key8342353
Date Received2019-02-15
Date Mfgr Received2018-06-07
Date Added to Maude2019-02-15
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag0
Product Problem Flag3
Reprocessed and Reused Flag3
Health Professional0
Initial Report to FDA0
Report to FDA3
Event Location3
Manufacturer ContactMR. MICHAEL SCHELLING
Manufacturer StreetJULIUS-HOELDER-STRASSE 36
Manufacturer CitySTUTTGART, 70597
Manufacturer CountryGM
Manufacturer Postal70597
Manufacturer G1STIHLER ELECTRONIC GMBH
Manufacturer StreetJULIUS-HOELER-STRASSE 36
Manufacturer CitySTUTTGART, 70597
Manufacturer CountryGM
Manufacturer Postal Code70597
Single Use3
Remedial ActionRC
Previous Use Code3
Event Type3
Type of Report0

Device Details

Brand NamePRISMAFLO II OR PRISMAFLO IIS
Generic NameBLOOD WARMER
Product CodeKOC
Date Received2019-02-15
Model NumberNO INFORMATION
Catalog NumberNO INFORMATION
Device AvailabilityN
Device Eval'ed by MfgrR
Device Sequence No1
Device Event Key0
ManufacturerSTIHLER ELECTRONIC GMBH
Manufacturer AddressJULIUS-HOELDER-STRASSE 36 STUTTGART, 70597 GM 70597


Patients

Patient NumberTreatmentOutcomeDate
10 2019-02-15