STATIC AND DYNAMIC COMPLIANCE PHYSICIAN CHOICE ENCORE TOWER 777543-102

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed from a health professional,user faci report with the FDA on 2019-02-15 for STATIC AND DYNAMIC COMPLIANCE PHYSICIAN CHOICE ENCORE TOWER 777543-102 manufactured by Vyaire Medical.

Event Text Entries

[136526756] (b)(4). The suspected device was received and a failure investigation was conducted. The reported issue was confirmed and a damaged (burnt) contact was found on the isolation transformer's switch. This issue will be internally investigated within vyaire.
Patient Sequence No: 1, Text Type: N, H10


[136526757] The customer initially reported the static and dynamic compliance vmax device was unable to power on. The customer in conjunction with the field service representative determined the most likely caused of event vmax cart. Customer reported there was no patient involvement with the event.
Patient Sequence No: 1, Text Type: D, B5


MAUDE Entry Details

Report Number2021710-2019-09890
MDR Report Key8345688
Report SourceHEALTH PROFESSIONAL,USER FACI
Date Received2019-02-15
Date of Report2019-02-07
Date of Event2018-08-22
Date Mfgr Received2019-01-31
Device Manufacturer Date2010-04-22
Date Added to Maude2019-02-15
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Reporter OccupationOTHER HEALTH CARE PROFESSIONAL
Health Professional3
Initial Report to FDA3
Report to FDA3
Event Location3
Manufacturer ContactSTANLEY TAN
Manufacturer Street22745 SAVI RANCH PKWY
Manufacturer CityYORBA LINDA CA 92887
Manufacturer CountryUS
Manufacturer Postal92887
Manufacturer Phone7149193324
Manufacturer G1VYAIRE MEDICAL INC.
Manufacturer Street1100 BIRD CENTER DRIVE
Manufacturer CityPALM SPRINGS CA 92262
Manufacturer CountryUS
Manufacturer Postal Code92262
Single Use3
Previous Use Code3
Event Type3
Type of Report3

Device Details

Brand NameSTATIC AND DYNAMIC COMPLIANCE
Generic NameCALCULATOR, PREDICATED VALUES, PULMONARY FUNCTION
Product CodeBTY
Date Received2019-02-15
Returned To Mfg2018-09-19
Model NumberPHYSICIAN CHOICE ENCORE TOWER
Catalog Number777543-102
OperatorHEALTH PROFESSIONAL
Device AvailabilityR
Device Eval'ed by MfgrY
Device Sequence No1
Device Event Key0
ManufacturerVYAIRE MEDICAL
Manufacturer Address22745 SAVI RANCH PKWY YORBA LINDA CA 92887 US 92887


Patients

Patient NumberTreatmentOutcomeDate
10 2019-02-15

© 2026 FDA.report