MILLENIUM HX

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed from a company representative,user f report with the FDA on 2019-02-15 for MILLENIUM HX manufactured by Mar Cor Purification.

Event Text Entries

[136457932] The facility reported observing a brown substance on the reverse osmosis (ro) membrane in their millenium hx portable ro system. There is potential for patient or user exposure. The facility immediately replaced the affected membrane in their ro system and has not experienced any issues. The source of the brown substance was confirmed to be the adhesive used in the construction of the membrane itself. It is unknown if the substance is entering the ro system fluid pathway. Additional testing and investigation is underway. There have been no reports of any patient adverse effects. This report is being submitted conservatively as the potential for patient or user exposure to the substance is unknown at this time. This complaint will continue being monitored in the mar cor purification/cantel complaint handling system.
Patient Sequence No: 1, Text Type: N, H10


[136457933] The facility reported observing a brown substance on the reverse osmosis (ro) membrane in their millenium hx portable ro system. There is potential for patient or user exposure.
Patient Sequence No: 1, Text Type: D, B5


MAUDE Entry Details

Report Number3019131-2019-00002
MDR Report Key8345765
Report SourceCOMPANY REPRESENTATIVE,USER F
Date Received2019-02-15
Date of Report2019-02-15
Date of Event2019-01-16
Date Mfgr Received2019-01-16
Device Manufacturer Date2015-10-23
Date Added to Maude2019-02-15
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Health Professional3
Initial Report to FDA3
Report to FDA3
Event Location3
Manufacturer ContactALEX NELSON
Manufacturer Street14550 28TH AVENUE NORTH
Manufacturer CityMINNEAPOLIS MN 55447
Manufacturer CountryUS
Manufacturer Postal55447
Manufacturer Phone7635094799
Manufacturer G1MAR COR PURIFICATION
Manufacturer Street14550 28TH AVENUE NORTH
Manufacturer CityMINNEAPOLIS MN 55447
Manufacturer CountryUS
Manufacturer Postal Code55447
Single Use3
Previous Use Code3
Event Type3
Type of Report3

Device Details

Brand NameMILLENIUM HX
Generic NameWATER PURIFICATION SYSTEM
Product CodeFIP
Date Received2019-02-15
Device AvailabilityY
Device Eval'ed by MfgrY
Device Sequence No1
Device Event Key0
ManufacturerMAR COR PURIFICATION
Manufacturer Address14550 28TH AVENUE NORTH MINNEAPOLIS MN 55447 US 55447


Patients

Patient NumberTreatmentOutcomeDate
101. Other 2019-02-15