PARADIGM REAL-TIME INSULIN INFUSION PUMP MMT-722LWWS

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed from a company representative,foreig report with the FDA on 2019-02-15 for PARADIGM REAL-TIME INSULIN INFUSION PUMP MMT-722LWWS manufactured by Medtronic Minimed.

Event Text Entries

[136459559] Currently it is unknown whether or not the device may have caused or contributed to the event as no product has been returned. The device will be returned for analysis and further information will follow once the analysis has been completed. No conclusion can be drawn at this time.
Patient Sequence No: 1, Text Type: N, H10


[136459560] It was reported that the customer's insulin pump had received motor error. The customer? S blood glucose level was 10 mmol/l. The customer reported that they received a motor error alarm. Customer was able to successfully clear the alarm. Customer was unable to complete pump rewind. The customer reported that the drive support cap was normal. The customer was advised the pump will need to be replaced. The customer was advised to discontinue use of the pump and was advised to refer to back up plan, per health care professional instructions. The insulin pump will be returned for analysis.
Patient Sequence No: 1, Text Type: D, B5


[141437934] Motor error alarm during rewind due to motor encoder signal out of phase. Unable to perform any tests due to motor error alarm.
Patient Sequence No: 1, Text Type: N, H10


MAUDE Entry Details

Report Number2032227-2019-01514
MDR Report Key8346014
Report SourceCOMPANY REPRESENTATIVE,FOREIG
Date Received2019-02-15
Date of Report2019-04-10
Date of Event2019-02-11
Date Mfgr Received2019-04-05
Device Manufacturer Date2014-03-26
Date Added to Maude2019-02-15
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Health Professional3
Initial Report to FDA3
Report to FDA3
Event Location3
Manufacturer ContactGERWIN DE GRAAFF
Manufacturer Street18000 DEVONSHIRE ST.
Manufacturer CityNORTHRIDGE CA 913251219
Manufacturer CountryUS
Manufacturer Postal913251219
Manufacturer Phone8185464805
Manufacturer G1MEDTRONIC MINIMED
Manufacturer Street18000 DEVONSHIRE ST.
Manufacturer CityNORTHRIDGE CA 913251219
Manufacturer CountryUS
Manufacturer Postal Code913251219
Single Use3
Previous Use Code3
Event Type3
Type of Report3

Device Details

Brand NamePARADIGM REAL-TIME INSULIN INFUSION PUMP MMT-722LWWS
Generic NamePUMP, INFUSION, INSULIN, TO BE USED WITH INVASIVE GLUCOSE SENSOR
Product CodeOYC
Date Received2019-02-15
Returned To Mfg2019-03-16
Model NumberMMT-722LWWS
Catalog NumberMMT-722LWWS
Lot NumberA000117169
OperatorLAY USER/PATIENT
Device AvailabilityR
Device AgeDA
Device Eval'ed by MfgrR
Device Sequence No1
Device Event Key0
ManufacturerMEDTRONIC MINIMED
Manufacturer Address18000 DEVONSHIRE ST. NORTHRIDGE CA 913251219 US 913251219


Patients

Patient NumberTreatmentOutcomeDate
10 2019-02-15

© 2025 FDA.report
This site is not affiliated with or endorsed by the FDA.