MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed from a company representative,foreig report with the FDA on 2019-02-18 for OPTIFLUX F18NRE DIALYZER FINISHED ASSY 0500308E manufactured by Ogden Manufacturing Plant.
[136865313]
The plant investigation is in process. A supplemental mdr will be submitted upon completion of this activity.
Patient Sequence No: 1, Text Type: N, H10
[136865314]
A user facility reported that a dialyzer blood leak occurred three minutes after the initiation of the patient? S hemodialysis (hd) treatment. It was noted that the blood leak was coming from the upper head of the dialyzer. The patient's estimated blood loss (ebl) was 20 ml. Upon follow up, it was confirmed that the machine, a fresenius 4008s machine, alarmed appropriately with a blood leak alarm. Blood leak test strips were not used. The leak was confirmed to be internal. There was no defect or damage seen on the dialyzer. There was no patient injury, adverse events, or medical intervention required as a result of the reported event. The patient was restarted on a new machine and treatment completed successfully with new supplies. The dialyzer was discarded and is not available to be returned to the manufacturer for physical evaluation
Patient Sequence No: 1, Text Type: D, B5
Report Number | 1713747-2019-00059 |
MDR Report Key | 8348051 |
Report Source | COMPANY REPRESENTATIVE,FOREIG |
Date Received | 2019-02-18 |
Date of Report | 2019-05-02 |
Date of Event | 2019-02-08 |
Date Mfgr Received | 2019-04-26 |
Device Manufacturer Date | 2018-05-29 |
Date Added to Maude | 2019-02-18 |
Event Key | 0 |
Report Source Code | Manufacturer report |
Manufacturer Link | Y |
Number of Patients in Event | 0 |
Adverse Event Flag | 3 |
Product Problem Flag | 3 |
Reprocessed and Reused Flag | 3 |
Health Professional | 3 |
Initial Report to FDA | 3 |
Report to FDA | 3 |
Event Location | 3 |
Manufacturer Contact | MATTHEW AMARAL |
Manufacturer Street | 920 WINTER ST. |
Manufacturer City | WALTHAM MA 02451 |
Manufacturer Country | US |
Manufacturer Postal | 02451 |
Manufacturer Phone | 7816999758 |
Manufacturer G1 | OGDEN MANUFACTURING PLANT |
Manufacturer Street | 475 WEST 13TH STREET |
Manufacturer City | OGDEN UT 84404 |
Manufacturer Country | US |
Manufacturer Postal Code | 84404 |
Single Use | 3 |
Previous Use Code | 3 |
Event Type | 3 |
Type of Report | 3 |
Brand Name | OPTIFLUX F18NRE DIALYZER FINISHED ASSY |
Generic Name | DIALYZER, CAPILLARY, HOLLOW FIBER |
Product Code | FJI |
Date Received | 2019-02-18 |
Catalog Number | 0500308E |
Lot Number | 18EU01016 |
Operator | HEALTH PROFESSIONAL |
Device Availability | N |
Device Age | MO |
Device Eval'ed by Mfgr | R |
Device Sequence No | 1 |
Device Event Key | 0 |
Manufacturer | OGDEN MANUFACTURING PLANT |
Manufacturer Address | 475 WEST 13TH STREET OGDEN UT 84404 US 84404 |
Patient Number | Treatment | Outcome | Date |
---|---|---|---|
1 | 0 | 2019-02-18 |