LIGHT SOURCE, 500XL, XENON 72200568

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed with the FDA on 2019-02-19 for LIGHT SOURCE, 500XL, XENON 72200568 manufactured by Smith & Nephew, Inc..

Event Text Entries

[136611780]
Patient Sequence No: 1, Text Type: N, H10


[136611781] It was reported that during procedure, device was not recognized. There was no delay reported and no back device available. No patient injury or other complications were reported.
Patient Sequence No: 1, Text Type: D, B5


[139320881] The device was received for evaluation. There was a relationship found between the returned device and the reported incident. A visual inspection was performed on the exterior of product and no physical damage was observed. Complaint of device not being recognized was confirmed. Light source failed power up. Cause of power failure is a defective power supply/ballast. Unit powered up and passed functional testing with a known good power supply installed. The complaint investigation has concluded the cause of the failure to be a defective electronic component. A review of the device history records showed there were no indications to suggest that the product did not meet manufacturing specification or would not be able to perform as intended.
Patient Sequence No: 1, Text Type: N, H10


MAUDE Entry Details

Report Number1643264-2019-00113
MDR Report Key8349431
Date Received2019-02-19
Date of Report2019-03-20
Date of Event2001-01-01
Date Mfgr Received2019-03-19
Date Added to Maude2019-02-19
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Reporter OccupationOTHER HEALTH CARE PROFESSIONAL
Health Professional3
Initial Report to FDA3
Report to FDA3
Event Location3
Manufacturer ContactJIM GONZALES
Manufacturer Street7000 WEST WILLIAM CANNON DRIVE
Manufacturer CityAUSTIN TX 78735
Manufacturer CountryUS
Manufacturer Postal78735
Manufacturer Phone5123585706
Manufacturer G1SMITH & NEPHEW, INC.
Manufacturer Street76 S. MERIDIAN AVE.
Manufacturer CityOKLAHOMA CITY OK 731076512
Manufacturer CountryUS
Manufacturer Postal Code731076512
Single Use3
Previous Use Code3
Event Type3
Type of Report0

Device Details

Brand NameLIGHT SOURCE, 500XL, XENON
Generic NameIMAGE, ILLUMINATION, FIBEROPTIC, FOR ENDOSCOPE
Product CodeFFS
Date Received2019-02-19
Returned To Mfg2019-02-06
Catalog Number72200568
OperatorHEALTH PROFESSIONAL
Device AvailabilityR
Device AgeDA
Device Eval'ed by MfgrR
Device Sequence No1
Device Event Key0
ManufacturerSMITH & NEPHEW, INC.
Manufacturer Address76 S. MERIDIAN AVE. OKLAHOMA CITY OK 731076512 US 731076512


Patients

Patient NumberTreatmentOutcomeDate
10 2019-02-19

© 2026 FDA.report
This site is not affiliated with or endorsed by the FDA.