[140013250] Sequencing results interpretation: the sample is homozygous at c. 48g>c as well as homozygous for the c-linked intronic marker which would be consistent with a c-c+ individual 1,2. However, after repeated attempts, exon 2 of the rhce gene was not able to be amplified, impeding the ability to make a definitive c/c determination. All exons for rhce were sequenced, no other polymorphisms were detected that would explain the c+ serological result. Sequencing could not confirm the precise type hea beadchip heah0225_g3 and heah2128_a2 results or the serological results. The most likely cause is a deletion, hybrid or polymorphic event affecting exon 2 of rhce. Hybrids, deletions and polymorphisms affecting expression are limitations of the precise type hea beadchip test. [(b)(4)].
Patient Sequence No: 1, Text Type: N, H10


[140013251] The customer reported a possible discrepancy. The donor is c- using the bioarray hea molecular beadchip kit; serology results were c+.
Patient Sequence No: 1, Text Type: D, B5


MAUDE Entry Details

Report Number3005967741-2019-00001
MDR Report Key8350592
Date Received2019-02-19
Date of Report2019-02-19
Date of Event2019-01-29
Date Mfgr Received2019-01-29
Device Manufacturer Date2018-11-13
Date Added to Maude2019-02-19
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Health Professional3
Initial Report to FDA3
Report to FDA3
Event Location3
Manufacturer ContactMS. LUZ VILLAMIZAR
Manufacturer Street35 TECHNOLOGY DRIVE SUITE 100
Manufacturer CityWARREN NJ 07059
Manufacturer CountryUS
Manufacturer Postal07059
Manufacturer Phone9084449591
Single Use3
Previous Use Code3
Event Type3
Type of Report0

Device Details

Brand NameHEA 1.2 BEADCHIP KIT, SLIDE
Generic NameHEA 1.2 BEADCHIP KIT, SLIDE
Product CodePEP
Date Received2019-02-19
Model Number800-20202-08
Catalog Number800-20202-08
Lot Number19-22-R-V / 19-85-V
Device AvailabilityY
Device Eval'ed by MfgrY
Device Sequence No1
Device Event Key0
ManufacturerBIOARRAY SOLUTIONS LTD.
Manufacturer Address35 TECHNOLOGY DRIVE SUITE 100 WARREN NJ 07059 US 07059


Patients

Patient NumberTreatmentOutcomeDate
10 2019-02-19

© 2026 FDA.report
This site is not affiliated with or endorsed by the FDA.