CLINICAL CHEMISTRY MAGNESIUM 07D70-21

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed from a health professional report with the FDA on 2019-02-19 for CLINICAL CHEMISTRY MAGNESIUM 07D70-21 manufactured by Abbott Manufacturing Inc.

Event Text Entries

[137141057] This issue was previously reported in manufacturers report 1628664-2018-02155, but it is unclear which suspect medical device lot number the customer was using; therefore, this report is being sent on the second lot. Further investigation of the customer issue included a review of the complaint text, a search for similar complaints, review of field data, and a review of labeling. No adverse trend was identified for the customer's issue. No return patient sample was available. Historical performance of the reagent lot was evaluated using world wide data. The patient data was analyzed and within established control limits. Labeling was reviewed and found to be adequate. Based on all available information and abbott diagnostics' complaint investigation no product deficiency was identified.
Patient Sequence No: 1, Text Type: N, H10


[137141058] The customer observed a falsely elevated magnesium result on the architect c4000 analyzer. The following data was provided (mg/dl): sid (b)(6) initial 9. 4, repeat 2. 1. Sid (b)(6) initial 6. 3, repeat 1. 6. There was no impact to patient management reported.
Patient Sequence No: 1, Text Type: D, B5


MAUDE Entry Details

Report Number1628664-2019-00145
MDR Report Key8351660
Report SourceHEALTH PROFESSIONAL
Date Received2019-02-19
Date of Report2019-02-19
Date Mfgr Received2019-02-15
Device Manufacturer Date2018-07-19
Date Added to Maude2019-02-19
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Reporter OccupationOTHER HEALTH CARE PROFESSIONAL
Health Professional3
Initial Report to FDA3
Report to FDA3
Event Location3
Manufacturer ContactNOEMI ROMERO-KONDOS, RN BSN
Manufacturer Street100 ABBOTT PARK ROAD DEPT. 09B9, LCCP1-3
Manufacturer CityABBOTT PARK IL 600643537
Manufacturer CountryUS
Manufacturer Postal600643537
Manufacturer Phone224667-512
Manufacturer G1ABBOTT MANUFACTURING INC
Manufacturer Street1921 HURD DRIVE
Manufacturer CityIRVING TX 75038
Manufacturer CountryUS
Manufacturer Postal Code75038
Single Use3
Previous Use Code3
Event Type3
Type of Report3

Device Details

Brand NameCLINICAL CHEMISTRY MAGNESIUM
Generic NameMAGNESIUM
Product CodeJGJ
Date Received2019-02-19
Catalog Number07D70-21
Lot Number33785UN18
OperatorHEALTH PROFESSIONAL
Device AvailabilityY
Device Eval'ed by MfgrY
Device Sequence No1
Device Event Key0
ManufacturerABBOTT MANUFACTURING INC
Manufacturer Address1921 HURD DRIVE IRVING TX 75038 US 75038


Patients

Patient NumberTreatmentOutcomeDate
10 2019-02-19

© 2024 FDA.report