CUSA EXCEL CONSOLE CUSAEXCEL

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed with the FDA on 2007-04-02 for CUSA EXCEL CONSOLE CUSAEXCEL manufactured by Valleylab.

Event Text Entries

[558482] The physician has alleged that the unit activated the ultrasonics on its own. As a result, the pt sustained a burn of the dura.
Patient Sequence No: 1, Text Type: D, B5


[7956689] The mfr has been notified of the reported complaint. The group leader of software support visited the user facility and tested each of the 4 hand pieces in the presence of the nurse who reported the incident. Each hand piece was observed without activation for 5 minutes, and then activated to ensure its proper function for 2 minutes. None of the hand pieces self activated during the testing. Subsequently, the nurse at the user facility acknowledged that the equipment is functioning properly. Based on the reported info and the onsite eval of the equipment, we are unable to confirm or determine the root cause of the alleged incident.
Patient Sequence No: 1, Text Type: N, H10


MAUDE Entry Details

Report Number3003418325-2007-00003
MDR Report Key836008
Date Received2007-04-02
Date of Report2007-03-30
Date Facility Aware2007-01-25
Report Date2007-02-27
Date Mfgr Received2007-01-25
Date Added to Maude2007-04-12
Event Key0
Report Source CodeDistributor report
Manufacturer LinkN
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Health Professional3
Initial Report to FDA3
Report to FDA3
Event Location0
Manufacturer ContactLINDA GOODSTEIN
Manufacturer Street311 ENTERPRISE DRIVE
Manufacturer CityPLAINSBORO NJ 08536
Manufacturer CountryUS
Manufacturer Postal08536
Manufacturer Phone6099362269
Manufacturer G1VALLEYLAB
Manufacturer Street5920 LONGBOW DRIVE BLDG 1, SERVICE DEPT
Manufacturer CityBOULDER CO 80301
Manufacturer CountryUS
Manufacturer Postal Code80301
Single Use3
Previous Use Code3
Event Type3
Type of Report3

Device Details

Brand NameCUSA EXCEL CONSOLE
Product CodeLBK
Date Received2007-04-02
Catalog NumberCUSAEXCEL
OperatorHEALTH PROFESSIONAL
Device AvailabilityN
Device AgeDA
Device Eval'ed by MfgrR
Implant FlagN
Date RemovedB
Device Sequence No1
Device Event Key823300
ManufacturerVALLEYLAB
Manufacturer AddressBOULDER CO US


Patients

Patient NumberTreatmentOutcomeDate
10 2007-04-02

© 2026 FDA.report
This site is not affiliated with or endorsed by the FDA.