PARKER BATH AL27110-EU

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed from a company representative,user f report with the FDA on 2019-02-22 for PARKER BATH AL27110-EU manufactured by Arjohuntleigh Polska Sp Z O.o..

Event Text Entries

[136855692] (b)(4). The device was evaluated by the arjo qualified representative. According to the results of inspection, the door lock mechanism was loose in such a way that the door lock lever has a double angle of up and down movement. If the lever was moved upwards, it was scraping off the protrusion on the door. Another problem is in the pin of the door, the door moves on the pin even when the pin is supported by a stop. The customer requested to exchange the whole device. The investigation is on-going and conclusion is not yet available. Additional information will be provided within the next report.
Patient Sequence No: 1, Text Type: N, H10


[136855693] It was reported that during use the bath's door lock disengaged, the door flipped out and made a mark on the wall. The customer stated that at the closing of the bathtub by the handle, there are symbols of "locked" and "unlocked" door. The handle always remained within these symbols. However, at this time the customer notice that the handle did not stop at the "locked" symbol, but was going further. Neither patient involvement nor injury occurrence were indicated.
Patient Sequence No: 1, Text Type: D, B5


MAUDE Entry Details

Report Number3007420694-2019-00031
MDR Report Key8361269
Report SourceCOMPANY REPRESENTATIVE,USER F
Date Received2019-02-22
Date of Report2019-07-05
Date Facility Aware2019-01-28
Report Date2019-07-05
Date Reported to FDA2019-07-05
Date Reported to Mfgr2019-07-05
Date Mfgr Received2019-06-07
Device Manufacturer Date2017-08-07
Date Added to Maude2019-02-22
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Health Professional3
Initial Report to FDA3
Report to FDA3
Event Location3
Manufacturer ContactMS. KINGA STOLINSKA
Manufacturer StreetKS. WAWRZYNIAKA 2
Manufacturer CityKOMORNIKI, 62-052
Manufacturer CountryPL
Manufacturer Postal62-052
Manufacturer G1ARJOHUNTLEIGH POLSKA SP Z O.O.
Manufacturer StreetKS. WAWRZYNIAKA 2
Manufacturer CityKOMORNIKI, 62-052
Manufacturer CountryPL
Manufacturer Postal Code62-052
Single Use3
Previous Use Code3
Event Type3
Type of Report3

Device Details

Brand NamePARKER BATH
Generic NameBATH, HYDRO-MASSAGE
Product CodeILJ
Date Received2019-02-22
Model NumberAL27110-EU
Device AvailabilityY
Device Age1 YR
Device Eval'ed by MfgrR
Device Sequence No1
Device Event Key0
ManufacturerARJOHUNTLEIGH POLSKA SP Z O.O.
Manufacturer AddressKS. WAWRZYNIAKA 2 KOMORNIKI, 62-052 PL 62-052


Patients

Patient NumberTreatmentOutcomeDate
101. Other 2019-02-22

© 2026 FDA.report