MULTI-DRUG ONE STEP MULTI-LINE 12 DRUG SCREEN TEST DEVICE (URINE) DOA-1125

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed from a foreign,health professional report with the FDA on 2019-02-22 for MULTI-DRUG ONE STEP MULTI-LINE 12 DRUG SCREEN TEST DEVICE (URINE) DOA-1125 manufactured by Alere San Diego, Inc..

Event Text Entries

[137561477] Investigation pending.
Patient Sequence No: 1, Text Type: N, H10


[137561478] (b)(6) 2018: (b)(6)-year old female patient was admitted to the emergency department after accidentally ingesting benzodiazepines. A urine sample was tested on the multi-drug one step 12 drug screen test device, and the result was negative. Activated charcoal was administered and monitored. A few hours later, testing on the urine sample was repeated on the multi-drug one step 12 drug screen test device and positive results were obtained for benzodiazepines and cocaine. The customer is not questioning the positive cocaine result as they caught the mother smoking a mixture containing cocaine in the patient's room. Flumazenil was administered, but customer states they do not know whether the treatment was provided before or after the positive bzo result, since the patient had three episodes of desaturation in the first 12 hours. No adverse outcomes reported. Although further information was requested, no further information was provided.
Patient Sequence No: 1, Text Type: D, B5


MAUDE Entry Details

Report Number2027969-2019-00027
MDR Report Key8363939
Report SourceFOREIGN,HEALTH PROFESSIONAL
Date Received2019-02-22
Date of Report2019-05-23
Date Mfgr Received2019-05-08
Date Added to Maude2019-02-22
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Reporter OccupationOTHER HEALTH CARE PROFESSIONAL
Health Professional3
Initial Report to FDA3
Report to FDA3
Event Location3
Manufacturer ContactYA-LING KING
Manufacturer Street9975 SUMMERS RIDGE RD.
Manufacturer CitySAN DIEGO CA 92121
Manufacturer CountryUS
Manufacturer Postal92121
Manufacturer Phone8588052084
Single Use3
Previous Use Code3
Event Type3
Type of Report3

Device Details

Brand NameMULTI-DRUG ONE STEP MULTI-LINE 12 DRUG SCREEN TEST DEVICE (URINE)
Generic NameDRUG SCREEN
Product CodeDIO
Date Received2019-02-22
Returned To Mfg2019-04-30
Model NumberDOA-1125
Lot NumberDOA8090358
OperatorHEALTH PROFESSIONAL
Device AvailabilityR
Device AgeDA
Device Eval'ed by MfgrY
Device Sequence No1
Device Event Key0
ManufacturerALERE SAN DIEGO, INC.
Manufacturer Address9975 SUMMERS RIDGE RD. SAN DIEGO CA 92121 US 92121


Patients

Patient NumberTreatmentOutcomeDate
10 2019-02-22