MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed with the FDA on 2019-02-25 for PAEDIATRIC FLEXI-LOCK PARALLEL CIRCUIT 96" WITH Y-PIECE LUER ELBOW MINI FILTER 038-02-612U manufactured by Flexicare Medical Dongguan Ltd..
[137384118]
Root cause: brittleness of tubing material. Corrective actions: increase level of plasticiser (eva), modification to mold tool to remove any sharp edges, increased qa/operator inspection.
Patient Sequence No: 1, Text Type: N, H10
[137384119]
Pediatric anesthesia circuit leak-checked prior to use. Anesthesia induced on pediatric pt. Circuit expanded and could not ventilate pt due to leak in circuit. Pt ventilated with ambu-bag and circuit replaced with new one. Original circuit found to have a crack in the tubing.
Patient Sequence No: 1, Text Type: D, B5
Report Number | 3006061749-2019-00001 |
MDR Report Key | 8365571 |
Date Received | 2019-02-25 |
Date of Report | 2019-02-25 |
Date of Event | 2019-01-15 |
Date Mfgr Received | 2019-01-15 |
Device Manufacturer Date | 2018-06-30 |
Date Added to Maude | 2019-02-25 |
Event Key | 0 |
Report Source Code | Manufacturer report |
Manufacturer Link | Y |
Number of Patients in Event | 0 |
Adverse Event Flag | 3 |
Product Problem Flag | 3 |
Reprocessed and Reused Flag | 3 |
Health Professional | 3 |
Initial Report to FDA | 3 |
Report to FDA | 3 |
Event Location | 3 |
Manufacturer Contact | MR RICHARD DOWNS |
Manufacturer Street | CYNON VALLEY BUSINESS PARK |
Manufacturer City | MOUNTAIN ASH, RCT CF454ER |
Manufacturer Country | UK |
Manufacturer Postal | CF45 4ER |
Manufacturer G1 | FLEXICARE MEDICAL DONGGUAN LTD. |
Manufacturer Street | NO. B-15 XICHENG IND ZONE 1 HENGLI TOWN |
Manufacturer City | DONGGUAN CITY, GUANGDONG 523460 |
Manufacturer Country | CH |
Manufacturer Postal Code | 523460 |
Single Use | 3 |
Remedial Action | MA |
Previous Use Code | 3 |
Event Type | 3 |
Type of Report | 0 |
Brand Name | PAEDIATRIC FLEXI-LOCK PARALLEL CIRCUIT 96" WITH Y-PIECE LUER ELBOW MINI FILTER |
Generic Name | ANESTHESIA BREATHING CIRCUIT - EXPANDABLE |
Product Code | OFP |
Date Received | 2019-02-25 |
Model Number | 038-02-612U |
Catalog Number | 038-02-612U |
Lot Number | 180601433 |
Operator | HEALTH PROFESSIONAL |
Device Availability | Y |
Device Eval'ed by Mfgr | R |
Device Sequence No | 1 |
Device Event Key | 0 |
Manufacturer | FLEXICARE MEDICAL DONGGUAN LTD. |
Manufacturer Address | NO. B-15 XICHENG IND ZONE 1 HENGLI TOWN DONGGUAN CITY, GUANGDONG 523460 CH 523460 |
Patient Number | Treatment | Outcome | Date |
---|---|---|---|
1 | 0 | 1. Required No Informationntervention | 2019-02-25 |