ADULT CPAP / HFOT SHIELDED SINGLE LIMB HEATED WIRE BREATHING SYSTEM 1.6M KIT 1 038-31-510U 038-31-501U

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed from a company representative report with the FDA on 2019-02-25 for ADULT CPAP / HFOT SHIELDED SINGLE LIMB HEATED WIRE BREATHING SYSTEM 1.6M KIT 1 038-31-510U 038-31-501U manufactured by Flexicare Medical Dongguan Ltd..

Event Text Entries

[138131044] Device not returned, so tested parts held in stock manufactured at a similar time. Some parts dimensionally incorrect / out of tolerance. Some parts exhibiting flash.
Patient Sequence No: 1, Text Type: N, H10


[138131065] 2 x water chambers part of the 038-31-501u circuit kits would not fill with water on a high flow nasal cannula setup. Water only came partway down the tubing on the first but would not fill the chamber. Complete setup exchanged. Second chamber the line would not even fill.
Patient Sequence No: 1, Text Type: D, B5


MAUDE Entry Details

Report Number3006061749-2019-00005
MDR Report Key8366568
Report SourceCOMPANY REPRESENTATIVE
Date Received2019-02-25
Date of Report2019-02-25
Date of Event2019-02-06
Date Mfgr Received2019-02-07
Date Added to Maude2019-02-25
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Reporter OccupationOTHER HEALTH CARE PROFESSIONAL
Health Professional3
Initial Report to FDA3
Report to FDA3
Event Location3
Manufacturer ContactMR RICHARD DOWNS
Manufacturer StreetCYNON VALLEY BUSINESS PARK
Manufacturer CityMOUNTAIN ASH, RCT CF454ER
Manufacturer CountryUK
Manufacturer PostalCF45 4ER
Manufacturer G1FLEXICARE MEDICAL DONGGUAN LTD.
Manufacturer StreetNO. B-15 XICHENG IND ZONE 1 HENGLI TOWN
Manufacturer CityDONGGUAN CITY, GUANGDONG 523460
Manufacturer CountryCH
Manufacturer Postal Code523460
Single Use3
Remedial ActionIN
Previous Use Code3
Event Type3
Type of Report3

Device Details

Brand NameADULT CPAP / HFOT SHIELDED SINGLE LIMB HEATED WIRE BREATHING SYSTEM 1.6M KIT 1
Generic NameSINGLE LIMB HEATED CIRCUIT
Product CodeBZO
Date Received2019-02-25
Model Number038-31-510U
Catalog Number038-31-501U
Lot Number180601434
OperatorHEALTH PROFESSIONAL
Device AvailabilityN
Device Eval'ed by MfgrR
Device Sequence No1
Device Event Key0
ManufacturerFLEXICARE MEDICAL DONGGUAN LTD.
Manufacturer AddressNO. B-15 XICHENG IND ZONE 1 HENGLI TOWN DONGGUAN CITY, GUANGDONG 523460 CH 523460


Patients

Patient NumberTreatmentOutcomeDate
101. Required No Informationntervention 2019-02-25

© 2026 FDA.report
This site is not affiliated with or endorsed by the FDA.