PULL THRU STUBBY BRUSH 100405

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed from a company representative,user f report with the FDA on 2019-02-25 for PULL THRU STUBBY BRUSH 100405 manufactured by Medivators.

Event Text Entries

[137113839] The facility reported pieces from a pull thru stubby brush breaking off inside an endoscope while staff pre-cleaned the endoscope. The pieces fell inside a patient during the subsequent patient procedure thus there is risk of cross-contamination. While performing an egd procedure with an olympus gif-hq190 series endoscope the physician detected pieces from a pull thru brush inside a patient when using the forceps in the biopsy channel. The physician then removed the pieces. It was reported the facility's endoscope technician had manually cleaned the endoscope, but they did not notice the pieces were missing from the pull thru stubby brush. The facility sent the endoscope to the manufacture for repair. Medivators has not been informed of their results. The pull thru brush was not returned for investigation. No adverse event reported. This will continue to be monitored in the medivators complaint handling system.
Patient Sequence No: 1, Text Type: N, H10


[137113840] The facility reported pieces from a pull thru stubby brush breaking off inside an endoscope while staff pre-cleaned the endoscope. The pieces fell inside a patient during the subsequent patient procedure thus there is risk of cross-contamination.
Patient Sequence No: 1, Text Type: D, B5


MAUDE Entry Details

Report Number1651395-2019-00002
MDR Report Key8368460
Report SourceCOMPANY REPRESENTATIVE,USER F
Date Received2019-02-25
Date of Report2019-02-25
Date of Event2019-01-30
Date Mfgr Received2019-01-30
Device Manufacturer Date2018-12-11
Date Added to Maude2019-02-25
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Health Professional3
Initial Report to FDA3
Report to FDA3
Event Location3
Manufacturer ContactLAUREN JOHNSON
Manufacturer Street14605 28TH AVE N
Manufacturer CityPLYMOUTH MN 55447
Manufacturer CountryUS
Manufacturer Postal55447
Manufacturer G1MEDIVATORS
Manufacturer Street3150 POLLOCK DRIVE
Manufacturer CityCONROE TX 77303
Manufacturer CountryUS
Manufacturer Postal Code77303
Single Use3
Remedial ActionPM
Previous Use Code3
Event Type3
Type of Report3

Device Details

Brand NamePULL THRU STUBBY BRUSH
Generic NameENDOSCOPE CLEANING BRUSH
Product CodeMNL
Date Received2019-02-25
Model Number100405
Lot Number944114
Device AvailabilityN
Device Eval'ed by MfgrR
Device Sequence No1
Device Event Key0
ManufacturerMEDIVATORS
Manufacturer Address3150 POLLOCK DRIVE CONROE TX 77303 US 77303


Patients

Patient NumberTreatmentOutcomeDate
101. Other 2019-02-25

© 2026 FDA.report
This site is not affiliated with or endorsed by the FDA.