UL PRO FUSED HEADLIGHT CABLE 9FT, 275CM 001388LX9

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed from a user facility report with the FDA on 2019-02-26 for UL PRO FUSED HEADLIGHT CABLE 9FT, 275CM 001388LX9 manufactured by Integra York, Pa Inc..

Event Text Entries

[139840373] The product was returned for evaluation. A visual inspection of the returned fiber optic cable found that several sections of the fiber optic cable bundle approximately 6 inches from the strain relief assembly at the headlight-end exhibited damage similar to being compressed by heavy equipment such as gurney. The fused bundle as observed at the lens in the connector at the mlx insertion-end is melted. There were no separated fibers observed in the evaluation as reported. When the cable-bundle is damaged through compression, the intensity of the light does not properly flow through the cable causing a build-up of heat in the connector at the mlx insertion point. This condition results in the fiber optic bundle melting as observed. The reported complaint is unconfirmed. The root cause of the damage to the fiber optic cable is considered user error. No manufacturing, workmanship, or material deficiency has been identified. Udi #: (b)(4).
Patient Sequence No: 1, Text Type: N, H10


[139840374] A customer reported that the ax2100bif cable has separated from the plug and showing fibers on the 001388lx9 ul pro fused headlight cable. There was no patient contact, injury, or surgery delay reported. Request for additional information has been sent.
Patient Sequence No: 1, Text Type: D, B5


MAUDE Entry Details

Report Number2523190-2019-00030
MDR Report Key8370071
Report SourceUSER FACILITY
Date Received2019-02-26
Date of Report2019-01-15
Date Mfgr Received2019-02-05
Device Manufacturer Date2018-11-11
Date Added to Maude2019-02-26
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Health Professional3
Initial Report to FDA3
Report to FDA3
Event Location3
Manufacturer ContactUSER KIMBERLY SHELLY
Manufacturer Street311 ENTERPRISE DRIVE
Manufacturer CityPLAINSBORO NJ 08536
Manufacturer CountryUS
Manufacturer Postal08536
Manufacturer Phone6099362393
Manufacturer G1INTEGRA YORK, PA INC.
Manufacturer Street589 DAVIES DRIVE
Manufacturer CityYORK PA 17402
Manufacturer CountryUS
Manufacturer Postal Code17402
Single Use3
Previous Use Code3
Event Type3
Type of Report3

Device Details

Brand NameUL PRO FUSED HEADLIGHT CABLE 9FT, 275CM
Generic NameCABLES
Product CodeFST
Date Received2019-02-26
Returned To Mfg2019-01-31
Catalog Number001388LX9
Lot Number10121812
OperatorHEALTH PROFESSIONAL
Device AvailabilityR
Device Eval'ed by MfgrY
Device Sequence No1
Device Event Key0
ManufacturerINTEGRA YORK, PA INC.
Manufacturer Address589 DAVIES DRIVE 589 DAVIES DRIVE YORK PA 17402 US 17402


Patients

Patient NumberTreatmentOutcomeDate
10 2019-02-26

© 2024 FDA.report
This site is not affiliated with or endorsed by the FDA.