BIRD HIGH FLOW BLENDER SERIES 03800A

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed with the FDA on 2019-02-27 for BIRD HIGH FLOW BLENDER SERIES 03800A manufactured by Vyaire Medical.

Event Text Entries

[137539737] Vyaire file identification: (b)(4). Any additional information received from the customer will be included in a follow-up report. The suspect device was returned to vyaire's factory service for evaluation and repair. Factory service was able to confirm the reported event and reported the device is out of calibration. After re-calibrating the device, the issue was resolved. Factory service returned the device back to specifications.
Patient Sequence No: 1, Text Type: N, H10


[137539738] The customer reported while using the high flow microblender; the fraction of inspired oxygen (fio2) is out of specifications at 30%. The customer reported messing with the knob, adjusting the seat, and now it is all out of specifications. The customer reported to not being trained to perform overhaul of the device and wanted to return the device to vyaire's factory service for evaluation/repair. The customer reported there is no patient involvement associated with the event.
Patient Sequence No: 1, Text Type: D, B5


MAUDE Entry Details

Report Number2021710-2019-09943
MDR Report Key8376007
Date Received2019-02-27
Date of Report2019-01-29
Date of Event2019-01-29
Date Mfgr Received2019-01-29
Device Manufacturer Date2018-02-04
Date Added to Maude2019-02-27
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Reporter OccupationOTHER HEALTH CARE PROFESSIONAL
Health Professional3
Initial Report to FDA3
Report to FDA3
Event Location3
Manufacturer ContactSTANLEY TAN
Manufacturer Street22745 SAVI RANCH PKWY
Manufacturer CityYORBA LINDA CA 92887
Manufacturer CountryUS
Manufacturer Postal92887
Manufacturer Phone7149193324
Manufacturer G1VYAIRE MEDICAL INC.
Manufacturer Street1100 BIRD CENTER DRIVE
Manufacturer CityPALM SPRINGS CA 92262
Manufacturer CountryUS
Manufacturer Postal Code92262
Single Use3
Previous Use Code3
Event Type3
Type of Report0

Device Details

Brand NameBIRD HIGH FLOW BLENDER SERIES
Generic NameMIXER, BREATHING GASES, ANESTHESIA INHALATION
Product CodeBZR
Date Received2019-02-27
Returned To Mfg2019-02-05
Model NumberBLENDER
Catalog Number03800A
OperatorHEALTH PROFESSIONAL
Device AvailabilityR
Device Eval'ed by MfgrY
Device Sequence No1
Device Event Key0
ManufacturerVYAIRE MEDICAL
Manufacturer Address22745 SAVI RANCH PKWY YORBA LINDA CA 92887 US 92887


Patients

Patient NumberTreatmentOutcomeDate
10 2019-02-27

© 2024 FDA.report