HIP GENERAL OFFSET-30? BROACH HANDLE - LEFT 01.10.10.131

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed with the FDA on 2019-02-28 for HIP GENERAL OFFSET-30? BROACH HANDLE - LEFT 01.10.10.131 manufactured by Medacta International Sa.

Event Text Entries

[139072318] Batch review performed on 25 february 2019: lot 1111995c: (b)(4) items manufactured and released on 10-apr-2018. No anomalies found related to the problem. To date, no other similar event has been reported on this lot.
Patient Sequence No: 1, Text Type: N, H10


[139072319] During the impaction the ball of the broach handle fell out from the passing hole and the spring was broken, no debris fell in the patient. There was a delay of max. Five minutes.
Patient Sequence No: 1, Text Type: D, B5


MAUDE Entry Details

Report Number3005180920-2019-00097
MDR Report Key8377334
Date Received2019-02-28
Date of Report2019-03-26
Date of Event2019-01-29
Date Mfgr Received2019-01-29
Device Manufacturer Date2018-04-10
Date Added to Maude2019-02-28
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Health Professional3
Initial Report to FDA3
Report to FDA3
Event Location3
Manufacturer ContactMR STEFANO BAJ
Manufacturer StreetSTRADA REGINA
Manufacturer CityCASTEL SAN PIETRO, SWITZERLAND 6874
Manufacturer CountrySZ
Manufacturer Postal6874
Manufacturer G1MEDACTA INTERNATIONAL SA
Manufacturer StreetSTRADA REGINA
Manufacturer CityCASTEL SAN PIETRO, SWITZERLAND 6874
Manufacturer CountrySZ
Manufacturer Postal Code6874
Single Use3
Previous Use Code3
Event Type3
Type of Report0

Device Details

Brand NameHIP GENERAL OFFSET-30? BROACH HANDLE - LEFT
Generic NameSURGICAL INSTRUMENT FOR HIP SURGERY
Product CodeLHX
Date Received2019-02-28
Returned To Mfg2019-02-13
Catalog Number01.10.10.131
Lot Number1111995C
OperatorHEALTH PROFESSIONAL
Device AvailabilityR
Device AgeDA
Device Eval'ed by MfgrY
Device Sequence No1
Device Event Key0
ManufacturerMEDACTA INTERNATIONAL SA
Manufacturer AddressSTRADA REGINA CASTEL SAN PIETRO, SWITZERLAND 6874 SZ 6874


Patients

Patient NumberTreatmentOutcomeDate
101. Other 2019-02-28