FREEDOM EVO 200 MCA 30020020

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed with the FDA on 2019-03-01 for FREEDOM EVO 200 MCA 30020020 manufactured by Tecan Schweiz Ag.

Event Text Entries

[137950582] A field service engineer was dispatched to evaluate and replace the gas springs. A review of preventive maintenance history indicated the last replacement was in february 2016. The springs are required to be replaced every 3 years and so were just coming due for replacement. The fse determined the left gas spring had no resistance or strength left. Both gas springs were replaced as they were due this month. The springs met their expected lifetime. This report, in an abundance of caution, is filed as prior reports of evo 200 front panels falling have cause more serious injuries for which mdrs were filed.
Patient Sequence No: 1, Text Type: N, H10


[137950583] Laboratory called to report that the left gas spring had failed and the front panel on an evo 200 instrument fell and was caught by the operator as it was coming down. No injuries occurred. The right hinge was able to secure the panel if held in a fully upright position. A field service engineer was dispatched to replace the gas springs.
Patient Sequence No: 1, Text Type: D, B5


MAUDE Entry Details

Report Number3003402518-2019-00002
MDR Report Key8381210
Date Received2019-03-01
Date of Report2019-03-01
Date of Event2019-02-04
Date Mfgr Received2019-02-04
Device Manufacturer Date2010-12-16
Date Added to Maude2019-03-01
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Reporter OccupationOTHER HEALTH CARE PROFESSIONAL
Health Professional3
Initial Report to FDA3
Report to FDA3
Event Location3
Manufacturer ContactMS. JILLIAN WALKER
Manufacturer StreetSEESTRASSE 103
Manufacturer CityMAENNEDORF, ZH 8708
Manufacturer CountrySZ
Manufacturer Postal8708
Manufacturer G1TECAN SCHWEIZ AG
Manufacturer StreetSEESTRASSE 103
Manufacturer CityMAENNEDORF, ZH 8708
Manufacturer CountrySZ
Manufacturer Postal Code8708
Single Use3
Previous Use Code3
Event Type3
Type of Report0

Device Details

Brand NameFREEDOM EVO 200 MCA
Generic NameSTATION, PIPETTING AND DILUTING, FOR CLINICAL USE
Product CodeJQW
Date Received2019-03-01
Model Number30020020
Device AvailabilityY
Device Eval'ed by MfgrY
Device Sequence No1
Device Event Key0
ManufacturerTECAN SCHWEIZ AG
Manufacturer AddressSEESTRASSE 103 MAENNEDORF, ZH 8708 SZ 8708


Patients

Patient NumberTreatmentOutcomeDate
10 2019-03-01

© 2026 FDA.report
This site is not affiliated with or endorsed by the FDA.