FSC

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed from a company representative,foreig report with the FDA on 2019-03-01 for FSC manufactured by Getinge Disinfection Ab.

Event Text Entries

[139605510] (b)(4). The issue is being investigated by manufacturing site.
Patient Sequence No: 1, Text Type: N, H10


[139605511] On (b)(6) 2019 we became aware of an issue with a loading equipment outside the machine - fsc (free standing conveyor). As it was stated, the end pin on the conveyor was stuck in "in-position" resulting in a situation when a trolley detached from it. Fortunately, the user noticed the situation and was able to stop a falling trolley before it touched the ground. No injuries were reported in relation to this event. Nevertheless, we decided to report this complaint based on the potential for adverse event if it was to reoccur.
Patient Sequence No: 1, Text Type: D, B5


MAUDE Entry Details

Report Number9616031-2019-00006
MDR Report Key8381663
Report SourceCOMPANY REPRESENTATIVE,FOREIG
Date Received2019-03-01
Date of Report2019-05-28
Date Facility Aware2019-02-18
Report Date2019-03-26
Date Reported to FDA2019-03-26
Date Mfgr Received2019-02-18
Device Manufacturer Date2017-02-13
Date Added to Maude2019-03-01
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Health Professional3
Initial Report to FDA3
Report to FDA3
Event Location3
Manufacturer ContactMR DENNIS GENITO
Manufacturer Street45 BARBOUR POND DRIVE
Manufacturer CityWAYNE NJ 07470
Manufacturer CountryUS
Manufacturer Postal07470
Manufacturer Phone9737097515
Manufacturer G1GETINGE DISINFECTION AB
Manufacturer StreetLJUNGADALSGATAN 11 VAXJO KRONOBERGS IAN [SE-07]
Manufacturer CityVAXJO, 35115
Manufacturer CountrySW
Manufacturer Postal Code35115
Single Use3
Remedial ActionRP
Previous Use Code3
Event Type3
Type of Report3

Device Details

Brand NameFSC
Generic NameFSC
Product CodePEC
Date Received2019-03-01
Device AvailabilityY
Device AgeDA
Device Eval'ed by MfgrR
Device Sequence No1
Device Event Key0
ManufacturerGETINGE DISINFECTION AB
Manufacturer AddressLJUNGADALSGATAN 11 VAXJO KRONOBERGS IAN [SE-07] VAXJO, 35115 SW 35115


Patients

Patient NumberTreatmentOutcomeDate
10 2019-03-01

© 2026 FDA.report
This site is not affiliated with or endorsed by the FDA.