BALKIN UP AND OVER CONTRALATERAL DESIGN FLEXOR CHECK FLO INTRODUCER REF# G10383

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed with the FDA on 2019-02-28 for BALKIN UP AND OVER CONTRALATERAL DESIGN FLEXOR CHECK FLO INTRODUCER REF# G10383 manufactured by Cook, Inc.

Event Text Entries

[137686034] Right popliteal and tibioperoneal trunk atherectomy procedure when surgical team noticed, the balkin 6 fr introducer tip looked unusual under fluoroscopy. Introducer withdrawn, noticed wire was advanced past coating. Case continued with new sets of wire. It was determined that there was no missing tip but the internal wire advanced past the coating. Product was returned to vendor for analysis; pending further information.
Patient Sequence No: 1, Text Type: D, B5


MAUDE Entry Details

Report NumberMW5084545
MDR Report Key8382019
Date Received2019-02-28
Date of Report2019-02-27
Date of Event2019-01-29
Date Added to Maude2019-03-01
Event Key0
Report Source CodeVoluntary report
Manufacturer LinkN
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Health Professional3
Initial Report to FDA3
Report to FDA3
Event Location3
Single Use3
Previous Use Code3
Event Type3
Type of Report3

Device Sequence Number: 1

Brand NameBALKIN UP AND OVER CONTRALATERAL DESIGN FLEXOR CHECK FLO INTRODUCER
Generic NameINTRODUCER, CATHETER
Product CodeDYB
Date Received2019-02-28
Returned To Mfg2019-02-08
Catalog NumberREF# G10383
Lot Number8511824
Device AvailabilityR
Device Eval'ed by MfgrI
Device Sequence No1
Device Event Key0
ManufacturerCOOK, INC

Device Sequence Number: 2

Brand NameZIPWIRE 260-CM HYDROPHILIC GUIDE WIRE
Generic NameCATHETER, INTRODUCER KIT
Product CodeOFC
Date Received2019-02-28
Returned To Mfg2019-02-08
Catalog NumberREF #M00146154B0
Lot Number10886778
Device AvailabilityR
Device Eval'ed by MfgrI
Device Sequence No2
Device Event Key0
ManufacturerBOSTON SCIENTIFIC CORP


Patients

Patient NumberTreatmentOutcomeDate
10 2019-02-28

© 2026 FDA.report