PAXGENE? BLOOD RNA TUBE 762165

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed with the FDA on 2019-03-01 for PAXGENE? BLOOD RNA TUBE 762165 manufactured by Becton, Dickinson And Company (bd).

Event Text Entries

[137939396] Date of event: unknown. The date received by manufacturer has been used for this field. There were multiple lot numbers reported to be involved. The information for each lot number is as follows: medical device lot #: 4265247. Medical device expiration date: 2016-03-31. Device manufacture date: 2014-09-22. Medical device lot #: 6014599. Medical device expiration date: 2017-07-31. Device manufacture date: 2016-01-14. (b)(6). A device evaluation is anticipated, but has not yet begun. Upon completion of the investigation, a supplemental report will be filed.
Patient Sequence No: 1, Text Type: N, H10


[137939397] It was reported that the paxgene? Blood rna tube bottoms broke during the extraction and after the first centrifugation, having been used to draw blood in 2015 and then stored in "-80? C" temperatures for approximately three years. Lot #'s 4265247 and 6014599 were reported to have been involved in this event, but it is unknown how many occurrences happened within each lot. As reported by the customer, translated from french to english, "we wanted to report problems with paxgene tubes, ref (b)(4). These tubes were used to draw in 2015, and then stored at -80? C until 2018. During extraction, the first centrifugation, the bottom of the tubes broke. The lot numbers are: lot 4265247 and lot 6014599".
Patient Sequence No: 1, Text Type: D, B5


MAUDE Entry Details

Report Number9617032-2019-00245
MDR Report Key8383590
Date Received2019-03-01
Date of Report2019-03-26
Date of Event2019-02-14
Date Mfgr Received2019-02-14
Date Added to Maude2019-03-01
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Reporter OccupationOTHER HEALTH CARE PROFESSIONAL
Health Professional3
Initial Report to FDA3
Report to FDA3
Event Location3
Manufacturer ContactBDX BRETT WILKO
Manufacturer Street9450 SOUTH STATE STREET
Manufacturer CitySANDY UT 84070
Manufacturer CountryUS
Manufacturer Postal84070
Manufacturer Phone8015652341
Manufacturer G1BECTON, DICKINSON AND COMPANY (BD)
Manufacturer StreetBELLIVER WAY BELLIVER INDUSTRIAL ESTATE
Manufacturer CityPLYMOUTH
Manufacturer CountryUS
Single Use3
Previous Use Code3
Event Type3
Type of Report0

Device Details

Brand NamePAXGENE? BLOOD RNA TUBE
Generic NameRNA PREANALYTICAL SYSTEMS
Product CodeNTW
Date Received2019-03-01
Catalog Number762165
Lot NumberSEE SECTION H.10.
Device Availability*
Device AgeDA
Device Eval'ed by MfgrR
Device Sequence No1
Device Event Key0
ManufacturerBECTON, DICKINSON AND COMPANY (BD)
Manufacturer AddressBELLIVER WAY BELLIVER INDUSTRIAL ESTATE PLYMOUTH US


Patients

Patient NumberTreatmentOutcomeDate
101. Other 2019-03-01

© 2026 FDA.report