BACT/ALERT? BPN 279019

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed from a foreign,health professional,u report with the FDA on 2019-03-01 for BACT/ALERT? BPN 279019 manufactured by Biomerieux Inc..

Event Text Entries

[138225049] A customer in (b)(6) reported a false negative result when using the bact/alert? Bpn bottle. The bottle (id (b)(4)) was loaded on (b)(6) 2019 and flagged as negative on (b)(6) 2019. The bottle was then unloaded (b)(6) 2019, and the cell indicator was observed as yellow. The graph indicated growth and finished over 4000. The bottle was subcultured and contained facultative anaerobic gram negative sphere bacilli. The local field service engineer (fse) had previously visited the site and checked the instrument condition and found no issue. There is no indication or report from the hospital or treating physician to biom? Rieux that the discrepant result led to any adverse event related to the patient's state of health. A biom? Rieux internal investigation will be initiated.
Patient Sequence No: 1, Text Type: D, B5


MAUDE Entry Details

Report Number3002769706-2019-00005
MDR Report Key8383610
Report SourceFOREIGN,HEALTH PROFESSIONAL,U
Date Received2019-03-01
Date of Report2019-10-03
Date of Event2019-01-23
Date Mfgr Received2019-09-06
Date Added to Maude2019-03-01
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Reporter OccupationOTHER HEALTH CARE PROFESSIONAL
Health Professional3
Initial Report to FDA3
Report to FDA3
Event Location3
Manufacturer ContactMS. DEBRA BROYLES
Manufacturer Street595 ANGLUM ROAD
Manufacturer CityHAZELWOOD MO 63042
Manufacturer CountryUS
Manufacturer Postal63042
Manufacturer G1BIOMERIEUX INC.
Manufacturer Street100 RODOLPHE STREET
Manufacturer CityDURHAM NC 27712
Manufacturer CountryUS
Manufacturer Postal Code27712
Single Use3
Previous Use Code3
Event Type3
Type of Report3

Device Details

Brand NameBACT/ALERT? BPN
Generic NameBACT/ALERT? BPN
Product CodeMZC
Date Received2019-03-01
Catalog Number279019
OperatorHEALTH PROFESSIONAL
Device AvailabilityN
Device Eval'ed by MfgrR
Device Sequence No1
Device Event Key0
ManufacturerBIOMERIEUX INC.
Manufacturer Address100 RODOLPHE STREET DURHAM NC 27712 US 27712


Patients

Patient NumberTreatmentOutcomeDate
10 2019-03-01