OPTISITE ARTERIAL CANNULAE OPTI20

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed with the FDA on 2019-03-02 for OPTISITE ARTERIAL CANNULAE OPTI20 manufactured by Edwards Lifesciences.

Event Text Entries

[138416120] Evaluation summary: as received, a small tear approximately 2mm long was found on the sealed pouch packaging of returned device (see figure 2). No other visual damage, contamination, or other abnormalities were found to the sealed sterile barrier pouch packaging and device. Photos attached by complaint handler appeared consistent with lab findings. Any packaging non-conformance that results in a breach of the sterile package (i. E. Pinhole, tear or seal issue), has the potential to result in a serious infection. In this case, there was no report of patient injury. Per the product evaluation, the report of "a small hole was found in the package of the optisite cannula" was confirmed. Engineering task has been opened for further investigation. A supplemental report will be submitted upon completion.
Patient Sequence No: 1, Text Type: N, H10


[138416121] It was reported by the warehouse in (b)(6) that a small hole was found in the package of the arterial cannulae during incoming inspection.
Patient Sequence No: 1, Text Type: D, B5


MAUDE Entry Details

Report Number3008500478-2019-00110
MDR Report Key8384344
Date Received2019-03-02
Date of Report2019-01-18
Date of Event2019-01-17
Date Mfgr Received2019-04-16
Device Manufacturer Date2018-12-14
Date Added to Maude2019-03-02
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Reporter OccupationOTHER HEALTH CARE PROFESSIONAL
Health Professional3
Initial Report to FDA3
Report to FDA3
Event Location3
Manufacturer ContactMS. AMRITHA SRINIVASAN
Manufacturer Street1 EDWARDS WAY MLE2
Manufacturer CityIRVINE CA 92614
Manufacturer CountryUS
Manufacturer Postal92614
Manufacturer Phone9492504062
Manufacturer G1EDWARDS LIFESCIENCES LLC
Manufacturer Street1 EDWARDS WAY
Manufacturer CityIRVINE CA 92614
Manufacturer CountryUS
Manufacturer Postal Code92614
Single Use3
Previous Use Code3
Event Type3
Type of Report0

Device Details

Brand NameOPTISITE ARTERIAL CANNULAE
Generic NameARTERIAL CANNULAE
Product CodeDQR
Date Received2019-03-02
Returned To Mfg2019-02-04
Model NumberOPTI20
Catalog NumberOPTI20
OperatorHEALTH PROFESSIONAL
Device AvailabilityR
Device AgeDA
Device Eval'ed by MfgrR
Device Sequence No1
Device Event Key0
ManufacturerEDWARDS LIFESCIENCES
Manufacturer AddressONE EDWARDS WAY IRVINE CA 92614 US 92614


Patients

Patient NumberTreatmentOutcomeDate
10 2019-03-02

© 2026 FDA.report
This site is not affiliated with or endorsed by the FDA.