METRX SYSTEM 9560702

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed from a company representative,foreig report with the FDA on 2019-03-07 for METRX SYSTEM 9560702 manufactured by Warsaw Orthopedics.

Event Text Entries

[138068346] Neither the device nor films of applicable imaging studies were returned to the manufacturer for evaluation. Therefore, we are unable to determine the definitive cause of the reported event. Although it is unknown if the device caused or contributed to the reported event or not, we are filing this report for notification purposes. If information is provided in the future, a supplemental report will be issued.
Patient Sequence No: 1, Text Type: N, H10


[138068347] It was reported that intra-op, heat burn to patient's back where connection of light source to light cord attaches. Radiance light source used with size 20 tubes. The protective covering was not slid over the light source to light cord connection. After first tlif was complete, burns were noticed on the patients back and identified to be caused by the light source/light cord connection. A vaseline gauze was immediately placed on the burn, and the covering slid over the light source to light cable connection.
Patient Sequence No: 1, Text Type: D, B5


MAUDE Entry Details

Report Number1030489-2019-00238
MDR Report Key8398212
Report SourceCOMPANY REPRESENTATIVE,FOREIG
Date Received2019-03-07
Date of Report2019-03-07
Date of Event2019-02-08
Date Mfgr Received2019-02-08
Device Manufacturer Date2017-11-14
Date Added to Maude2019-03-07
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Health Professional3
Initial Report to FDA3
Report to FDA3
Event Location3
Manufacturer ContactSTACIE ZIEMBA
Manufacturer Street1800 PYRAMID PLACE
Manufacturer CityMEMPHIS TN 38132
Manufacturer CountryUS
Manufacturer Postal38132
Manufacturer Phone9013963133
Manufacturer G1WARSAW ORTHOPEDICS
Manufacturer Street2500 SILVEUS CROSSING
Manufacturer CityWARSAW IN 46582
Manufacturer CountryUS
Manufacturer Postal Code46582
Single Use3
Previous Use Code3
Event Type3
Type of Report3

Device Details

Brand NameMETRX SYSTEM
Generic NameLIGHT, SURGICAL, CARRIER
Product CodeFSZ
Date Received2019-03-07
Model NumberNA
Catalog Number9560702
Lot Number0601680W
OperatorHEALTH PROFESSIONAL
Device AvailabilityN
Device Eval'ed by MfgrN
Device Sequence No1
Device Event Key0
ManufacturerWARSAW ORTHOPEDICS
Manufacturer Address2500 SILVEUS CROSSING WARSAW IN 46582 US 46582


Patients

Patient NumberTreatmentOutcomeDate
101. Other 2019-03-07

© 2024 FDA.report