[138123699] The reasons for coming off the screws is unknown. Mechanical vibration or transportation of cart may have caused random loosening of the screws. Biomed at health care facility reinstall the camera just replacing the screws holding it. Biomed verified the camera functionality and safety check performed. The device is placed back in use and not returning for further investigation.
Patient Sequence No: 1, Text Type: N, H10


[138123700] Mounting screws were dislodged. Ipela dropped from the bracket. There is no injury to patient or user.
Patient Sequence No: 1, Text Type: D, B5


MAUDE Entry Details

Report Number9612330-2019-00001
MDR Report Key8399967
Report SourceHEALTH PROFESSIONAL
Date Received2019-03-07
Date of Report2019-02-06
Date of Event2019-02-06
Date Mfgr Received2019-02-06
Date Added to Maude2019-03-07
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Reporter OccupationOTHER HEALTH CARE PROFESSIONAL
Health Professional3
Initial Report to FDA3
Report to FDA3
Event Location3
Manufacturer ContactMR SANJAY MEHTA
Manufacturer Street2568 BRISTOL CIRCLE
Manufacturer CityOAKVILLE, ONTARIO L6H5S1
Manufacturer CountryCA
Manufacturer PostalL6H5S1
Manufacturer G1NATUS MEDICAL INCORPORATED DBA EXCEL-TECH (XLTEK)
Manufacturer Street2568 BRISTOL CIRCLE
Manufacturer CityOAKVILLE, ONTARIO L6H5S1
Manufacturer CountryCA
Manufacturer Postal CodeL6H5S1
Single Use3
Remedial ActionRP
Previous Use Code3
Event Type3
Type of Report3

Device Details

Brand NameERGOJUST CART
Generic NameELECTROENCEPHALOGRAPH
Product CodeGWQ
Date Received2019-03-07
Model NumberIP-EC-DT-CART
Catalog NumberN/A
Lot NumberN/A
OperatorHEALTH PROFESSIONAL
Device AvailabilityN
Device Eval'ed by MfgrR
Device Sequence No1
Device Event Key0
ManufacturerNATUS MEDICAL INC. DBA EXCLE-TECH LTD (XLTEK)
Manufacturer Address2568 BRISTOL CIRCLE OAKVILLE, OAKVILLE ONTARIO L6H 5S1 CA L6H 5S1


Patients

Patient NumberTreatmentOutcomeDate
101. Other 2019-03-07

© 2026 FDA.report
This site is not affiliated with or endorsed by the FDA.