INTEGRATED DIGITAL SURGICAL RACK

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed with the FDA on 2019-03-07 for INTEGRATED DIGITAL SURGICAL RACK manufactured by Black Diamond Video.

Event Text Entries

[138374309] User facility personnel immediately powered down the integrated digital surgical rack. A steris service technician arrived onsite to inspect the unit and found that the isolation transformer had overloaded. Isolation transformers are used to protect equipment from electrical surges. The technician replaced the isolation transformer, tested the unit, confirmed it to be operating according to specification, and returned it to service. A 3-year complaint review indicates this to be an isolated event. No additional issues have been reported.
Patient Sequence No: 1, Text Type: N, H10


[138374310] The user facility reported a burning odor and smoke emitting from their integrated digital surgical rack during a patient procedure. Report of staff nose and eye irritation. The procedure was completed successfully.
Patient Sequence No: 1, Text Type: D, B5


MAUDE Entry Details

Report Number3008776287-2019-00001
MDR Report Key8401729
Date Received2019-03-07
Date of Report2019-03-07
Date of Event2019-02-04
Date Mfgr Received2019-02-05
Date Added to Maude2019-03-07
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag0
Product Problem Flag3
Reprocessed and Reused Flag3
Reporter OccupationOTHER HEALTH CARE PROFESSIONAL
Health Professional3
Initial Report to FDA3
Report to FDA3
Event Location3
Manufacturer ContactMR. DANIEL DAVY
Manufacturer Street5960 HEISLEY ROAD
Manufacturer CityMENTOR OH 44060
Manufacturer CountryUS
Manufacturer Postal44060
Manufacturer Phone4403927453
Manufacturer G1BLACK DIAMOND VIDEO
Manufacturer Street503 CANAL BLVD.
Manufacturer CityRICHMOND CA 94804
Manufacturer CountryUS
Manufacturer Postal Code94804
Single Use3
Previous Use Code3
Event Type3
Type of Report0

Device Details

Brand NameINTEGRATED DIGITAL SURGICAL RACK
Generic NameINTEGRATED DIGITAL SURGICAL RACK
Product CodeFWC
Date Received2019-03-07
OperatorHEALTH PROFESSIONAL
Device AvailabilityY
Device Eval'ed by MfgrY
Device Sequence No1
Device Event Key0
ManufacturerBLACK DIAMOND VIDEO
Manufacturer Address503 CANAL BLVD. RICHMOND CA 94804 US 94804


Patients

Patient NumberTreatmentOutcomeDate
10 2019-03-07

© 2026 FDA.report