CALIPER IN & OUT DIA TO 150MM AA846R

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed with the FDA on 2019-03-08 for CALIPER IN & OUT DIA TO 150MM AA846R manufactured by Aesculap Ag.

Event Text Entries

[138527322] (b)(4). The two calipers are in a mint condition, but two adjustments screws and one guiding screw are not in place. Only one of the adjustment screws was provided, the other adjustment screw and the guiding screw are missing. Vigilance investigator carried out the pictorial documentation visually and microscopically. A caliper "a" one of the adjustment screws is not in place. At caliper "b" one of the adjustment screws and guiding screw are not in place. It is no longer traceable from to which caliper the provided screw belongs to. A review of the device quality and manufacturing history records was not possible because the lot number is unknown. Based on the information available as well as a result of the investigation the root cause of the failure is most probably related to an insufficient usage.
Patient Sequence No: 1, Text Type: N, H10


[138527323] It was reported by the healthcare professional "the screw was loose and this made is impossible to use the caliper. "
Patient Sequence No: 1, Text Type: D, B5


MAUDE Entry Details

Report Number9610612-2019-00171
MDR Report Key8404048
Date Received2019-03-08
Date of Report2019-03-07
Date Facility Aware2019-03-07
Date Mfgr Received2019-02-07
Date Added to Maude2019-03-08
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Reporter OccupationOTHER HEALTH CARE PROFESSIONAL
Health Professional3
Initial Report to FDA3
Report to FDA3
Event Location3
Manufacturer ContactMRS. NICOLE BROYLES
Manufacturer Street615 LAMBERT POINTE DRIVE
Manufacturer CityHAZELWOOD MO 63042
Manufacturer CountryUS
Manufacturer Postal63042
Manufacturer G1AESCULAP AG
Manufacturer StreetPO BOX 40
Manufacturer CityTUTTLINGEN, 78501
Manufacturer CountryGM
Manufacturer Postal Code78501
Single Use3
Previous Use Code3
Event Type3
Type of Report0

Device Details

Brand NameCALIPER IN & OUT DIA TO 150MM
Generic NameGENERAL SURGICAL INSTRUMENTS
Product CodeHOE
Date Received2019-03-08
Returned To Mfg2019-03-06
Model NumberAA846R
Catalog NumberAA846R
OperatorHEALTH PROFESSIONAL
Device AvailabilityR
Device Eval'ed by MfgrY
Device Sequence No1
Device Event Key0
ManufacturerAESCULAP AG
Manufacturer AddressPO BOX 40 TUTTLINGEN, 78501 GM 78501


Patients

Patient NumberTreatmentOutcomeDate
10 2019-03-08

© 2026 FDA.report
This site is not affiliated with or endorsed by the FDA.