TRIVEX SYSTEM RESECTOR HANDPIECE 7210387F

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed with the FDA on 2019-03-08 for TRIVEX SYSTEM RESECTOR HANDPIECE 7210387F manufactured by Lemaitre Vascular, Inc..

Event Text Entries

[139344604] We have received the complaint device for evaluation and we have confirmed the reported incident. The external housing and power cords were visually inspected and were found to be acceptable. When the handpiece was connected to the control unit, the drive shaft immediately started to rotate. The motor on the handpiece automatically engaged in the run mode and was over riding signals from either buttons if pressed. Further inspection found water and particulate inside the core tube. The likely root cause of the issue is a seal failure that led to water ingress into the handpiece, which adversely affected the motor functionality. We currently have a capa open to address this issue. The corrective action includes repairs by replacing the current seal housing with a new seal housing assembly, o-rings and switch pcb and motor if there is any evidence of water inside the core tube. We believe these changes will better prevent water and steam from entering into the inner components of the handpiece. Device was not used for the procedure. Procedure was completed using another handpiece that they had in stock.
Patient Sequence No: 1, Text Type: N, H10


[139344605] During pre-use check, the resector kept spinning even after releasing the button.
Patient Sequence No: 1, Text Type: D, B5


MAUDE Entry Details

Report Number1220948-2019-00023
MDR Report Key8405616
Date Received2019-03-08
Date of Report2019-03-08
Date of Event2019-02-08
Date Mfgr Received2019-02-08
Date Added to Maude2019-03-08
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Health Professional3
Initial Report to FDA3
Report to FDA3
Event Location3
Manufacturer ContactMR. PRAGYA THIKEY
Manufacturer Street63 SECOND AVE
Manufacturer CityBURLINGTON MA 01803
Manufacturer CountryUS
Manufacturer Postal01803
Manufacturer Phone7812212266
Manufacturer G1LEMAITRE VASCULAR, INC.
Manufacturer Street63 SECOND AVE
Manufacturer CityBURLINGTON MA 01803
Manufacturer CountryUS
Manufacturer Postal Code01803
Single Use3
Previous Use Code3
Event Type3
Type of Report0

Device Details

Brand NameTRIVEX SYSTEM RESECTOR HANDPIECE
Generic NameVARICOSE VEIN ALBATION SYSTEM
Product CodeDWQ
Date Received2019-03-08
Returned To Mfg2019-02-18
Catalog Number7210387F
OperatorHEALTH PROFESSIONAL
Device AvailabilityR
Device Eval'ed by MfgrY
Device Sequence No1
Device Event Key0
ManufacturerLEMAITRE VASCULAR, INC.
Manufacturer Address63 SECOND AVE BURLINGTON MA 02148 US 02148


Patients

Patient NumberTreatmentOutcomeDate
10 2019-03-08

© 2026 FDA.report
This site is not affiliated with or endorsed by the FDA.