STA COMPACT MAX PERCAGE 58989

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed from a foreign,health professional report with the FDA on 2019-03-11 for STA COMPACT MAX PERCAGE 58989 manufactured by Diagnostica Stago S.a.s..

Event Text Entries

[138428074] (b)(4). The event occurred in (b)(6) but the device is also marketed in the usa. The analyzer's logs have been retrieved. Investigations are still on going under corporate complaint (b)(4). Stago will provide a follow-up report(s) once relevant information has become available.
Patient Sequence No: 1, Text Type: N, H10


[138428075] A patient sample yielded a result of 215ng/ml utilizing sta liatest ddi plus (prod# 00662 - lot# 254415) on an sta compact max analyzer on (b)(6) 2019. The laboratory was then informed that the patient had a thrombosis: the clinician contacted the laboratory by phone asking if the result was correct since the patient showed clear symptoms of dvt. This was confirmed by the radiology department. The same tube was retested the next day and yielded a result of >4000ng/ml.
Patient Sequence No: 1, Text Type: D, B5


MAUDE Entry Details

Report Number8043273-2019-00002
MDR Report Key8408876
Report SourceFOREIGN,HEALTH PROFESSIONAL
Date Received2019-03-11
Date of Report2019-04-12
Date of Event2019-02-12
Date Facility Aware2019-03-06
Report Date2019-03-11
Date Reported to FDA2019-03-11
Date Reported to Mfgr2019-03-11
Date Mfgr Received2019-02-12
Device Manufacturer Date2017-10-02
Date Added to Maude2019-03-11
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Reporter OccupationOTHER HEALTH CARE PROFESSIONAL
Health Professional3
Initial Report to FDA3
Report to FDA3
Event Location3
Manufacturer ContactMR. WALID BEN AMMAR
Manufacturer Street2 RUE PIERRE FOSSATI
Manufacturer CityFRANCONVILLE VAL D'OISE, 95130
Manufacturer CountryFR
Manufacturer Postal95130
Manufacturer G1DIAGNOSTICA STAGO S.A.S.
Manufacturer Street3 ALLEE THERESA
Manufacturer CityASNIERES-SUR-SEINE, 92600
Manufacturer CountryFR
Manufacturer Postal Code92600
Single Use3
Previous Use Code3
Event Type3
Type of Report3

Device Details

Brand NameSTA COMPACT MAX PERCAGE
Generic NameIVD COAGULATION DEVICE/INSTRUMENT
Product CodeJPA
Date Received2019-03-11
Catalog Number58989
OperatorHEALTH PROFESSIONAL
Device AvailabilityY
Device Age1 YR
Device Eval'ed by MfgrN
Device Sequence No1
Device Event Key0
ManufacturerDIAGNOSTICA STAGO S.A.S.
Manufacturer Address3 ALL?E THERESA ASNI?RES-SUR-SEINE, 92600 FR 92600


Patients

Patient NumberTreatmentOutcomeDate
101. Other 2019-03-11