AUTOCLAVABLE INTERNAL HANDLES (MSERIES) 8011-0501-01 8011050101

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed with the FDA on 2019-03-11 for AUTOCLAVABLE INTERNAL HANDLES (MSERIES) 8011-0501-01 8011050101 manufactured by Zoll Medical Corporation.

Event Text Entries

[138432191] Zoll medical corporation has received the product and will be providing a supplemental report when our investigation is completed.
Patient Sequence No: 1, Text Type: N, H10


[138432192] Complainant alleged that during biomed testing, the associated defibrillator failed leakage current test using these internal handles. Complainant indicated that there was no patient involvement in the reported malfunction.
Patient Sequence No: 1, Text Type: D, B5


[139553956] The internal handles were returned to zoll medical corporation and the customer's report was not replicated or confirmed. The internal handles were put through extensive testing including full functionality testing multiple discharges and leakage testing without duplicating the report. The internal handles were returned to the customer. No trend is associated with reports of this type.
Patient Sequence No: 1, Text Type: N, H10


[145780603] After further investigation it was determined that the cause of the problem was due to a test set up configuration and was not due to a malfunction of the device. Analysis of reports of this type has not identified an increase in trend.
Patient Sequence No: 1, Text Type: N, H10


MAUDE Entry Details

Report Number1220908-2019-00584
MDR Report Key8409247
Date Received2019-03-11
Date of Report2019-02-18
Date Mfgr Received2019-02-18
Device Manufacturer Date2006-10-01
Date Added to Maude2019-03-11
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Health Professional3
Initial Report to FDA3
Report to FDA3
Event Location3
Manufacturer Street269 MILL ROAD
Manufacturer CityCHELMSFORD MA 01824
Manufacturer CountryUS
Manufacturer Postal01824
Manufacturer Phone9784219552
Single Use3
Previous Use Code3
Event Type3
Type of Report3

Device Details

Brand NameAUTOCLAVABLE INTERNAL HANDLES (MSERIES)
Generic NameINTERNAL PADDLE
Product CodeLDD
Date Received2019-03-11
Returned To Mfg2019-02-26
Model Number8011-0501-01
Catalog Number8011050101
Lot NumberNA
Device AvailabilityR
Device AgeDA
Device Eval'ed by MfgrY
Device Sequence No1
Device Event Key0
ManufacturerZOLL MEDICAL CORPORATION
Manufacturer Address269 MILL ROAD CHELMSFORD MA 01824 US 01824