VITROS CHEMISTRY PRODUCTS VALP REAGENT 6801710

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed with the FDA on 2019-03-11 for VITROS CHEMISTRY PRODUCTS VALP REAGENT 6801710 manufactured by Ortho-clinical Diagnostics.

Event Text Entries

[139328053] The investigation determined that higher and lower than expected vitros valproic acid (valp) results were obtained from non-vitros quality control (qc) fluids processed using vitros chemistry products valp reagent in combination with a vitros 5600 integrated system. A definitive assignable cause for the events could not be determined. Historical vitros valp quality control results confirm the observed imprecision, however the cause of the imprecision is unknown. The customer is calibrating frequently which may be contributing to the imprecision. The within run valp precision test results were not performed as requested, therefore a vitros 5600 integrated system issue could not be entirely ruled out. The vitros valp quality control results were acceptable after a recalibration event on (b)(6) 2019, therefore vitros valp lot 2511-26-6892 is not likely to be a contributor to the event. In addition, continual tracking and trending of complaints has not identified any signals that would point to a potential systemic issue with vitros valp reagent lot 2511-26-6892. No information was obtained regarding sample handling, it is possible the customer is not handling the control fluids consistently day to day which could contribute to imprecise results.
Patient Sequence No: 1, Text Type: N, H10


[139328054] A customer initially reported higher than expected vitros valproic acid (valp) quality results obtained from non-vitros quality control (qc) fluids using vitros chemistry products valp reagent on a vitros 5600 integrated system. An investigation later discovered lower than expected vitros valp results obtained from non-vitros quality control (qc) fluids using vitros chemistry products valp reagent on a vitros 5600 integrated system. Biorad immunoassay plus control lot 40951 vitros valp results 19. 5, 51. 4 and 52. 7 ug/ml versus the expected vitros valp result 40. 4 ug/ml. Biorad immunoassay plus control lot 40952 vitros valp results 52. 6, 91. 8 and 92. 5 ug/ml versus the expected vitros valp result 75. 8 ug/ml. Biorad immunoassay plus control lot 40953 vitros valp results 69. 9, 70. 8, 72. 0 and 78. 9 ug/ml versus the expected vitros valp result 109. 0 ug/ml. Biased results of the magnitude and direction observed may lead to inappropriate physician action if they were to occur undetected on patient samples. The higher and lower than expected vitros valp results were obtained when processing quality control fluids. However, the investigation cannot definitively conclude that patient sample results were not affected or would not be affected if the event were to recur undetected. There have been no allegations of patient harm as a result of this event. This report is number 1 of 2 mdr? S for this event. Two (2) 3500a forms are being submitted for this event as 2 devices were involved. This report corresponds to ortho clinical diagnostics inc. Complaint number (b)(4).
Patient Sequence No: 1, Text Type: D, B5


MAUDE Entry Details

Report Number1319808-2019-00010
MDR Report Key8410015
Date Received2019-03-11
Date of Report2019-03-11
Date of Event2019-02-06
Date Mfgr Received2019-02-14
Device Manufacturer Date2018-05-31
Date Added to Maude2019-03-11
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Health Professional3
Initial Report to FDA3
Report to FDA3
Event Location3
Manufacturer ContactMR. JAMES A STEVENS
Manufacturer Street100 INDIGO CREEK DRIVE
Manufacturer CityROCHESTER NY 14626
Manufacturer CountryUS
Manufacturer Postal14626
Manufacturer Phone5854533000
Manufacturer G1ORTHO-CLINICAL DIAGNOSTICS
Manufacturer Street1000 LEE ROAD
Manufacturer CityROCHESTER NY 14606
Manufacturer CountryUS
Manufacturer Postal Code14606
Single Use3
Previous Use Code3
Event Type3
Type of Report0

Device Details

Brand NameVITROS CHEMISTRY PRODUCTS VALP REAGENT
Generic NameIN VITRO DIAGNOSTICS
Product CodeLEG
Date Received2019-03-11
Catalog Number6801710
Lot Number2511-26-6892
OperatorHEALTH PROFESSIONAL
Device AvailabilityY
Device Eval'ed by MfgrY
Device Sequence No1
Device Event Key0
ManufacturerORTHO-CLINICAL DIAGNOSTICS
Manufacturer Address100 INDIGO CREEK DRIVE ROCHESTER NY 14626 US 14626


Patients

Patient NumberTreatmentOutcomeDate
10 2019-03-11

© 2024 FDA.report
This site is not affiliated with or endorsed by the FDA.