PAEDIATRIC PARALLEL CIRCUIT 96" Y-PIECE LUER ELBOW 2 X B/VIRAL FILTERS 1L R BAG 038-02-713U

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed from a health professional report with the FDA on 2019-03-12 for PAEDIATRIC PARALLEL CIRCUIT 96" Y-PIECE LUER ELBOW 2 X B/VIRAL FILTERS 1L R BAG 038-02-713U manufactured by Flexicare Medical Dongguan Ltd..

Event Text Entries

[138546521] Root cause: brittleness of tubing material corrective actions: increase level of plasticiser (eva), modification to mold tool to remove any sharp edges, increased qa/operator inspection.
Patient Sequence No: 1, Text Type: N, H10


[138546522] The seam on the side of the tubing failed and cause of failure to be able to ventilate ab anesthetized patient.
Patient Sequence No: 1, Text Type: D, B5


MAUDE Entry Details

Report Number3006061749-2019-00006
MDR Report Key8412557
Report SourceHEALTH PROFESSIONAL
Date Received2019-03-12
Date of Report2019-03-12
Date of Event2019-01-31
Date Mfgr Received2019-03-04
Device Manufacturer Date2018-07-31
Date Added to Maude2019-03-12
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Health Professional3
Initial Report to FDA3
Report to FDA3
Event Location3
Manufacturer ContactMR RICHARD DOWNS
Manufacturer StreetCYNON VALLEY BUSINESS PARK
Manufacturer CityMOUNTAIN ASH, RCT CF454ER
Manufacturer CountryUK
Manufacturer PostalCF45 4ER
Manufacturer G1FLEXICARE MEDICAL DONGGUAN LTD.
Manufacturer StreetNO. B-15 XICHENG IND ZONE 1 HENGLI TOWN
Manufacturer CityDONGGUAN CITY, GUANGDONG 523460
Manufacturer CountryCH
Manufacturer Postal Code523460
Single Use3
Remedial ActionMA
Previous Use Code3
Event Type3
Type of Report3

Device Details

Brand NamePAEDIATRIC PARALLEL CIRCUIT 96" Y-PIECE LUER ELBOW 2 X B/VIRAL FILTERS 1L R BAG
Generic NamePAEDIATRIC ANAESTHESIA CIRCUIT
Product CodeOFP
Date Received2019-03-12
Model Number038-02-713U
Catalog Number038-02-713U
Lot Number180701295
OperatorHEALTH PROFESSIONAL
Device AvailabilityY
Device Eval'ed by MfgrN
Device Sequence No1
Device Event Key0
ManufacturerFLEXICARE MEDICAL DONGGUAN LTD.
Manufacturer AddressNO. B-15 XICHENG IND ZONE 1 HENGLI TOWN DONGGUAN CITY, GUANGDONG 523460 CH 523460


Patients

Patient NumberTreatmentOutcomeDate
101. Required No Informationntervention 2019-03-12

© 2026 FDA.report
This site is not affiliated with or endorsed by the FDA.