Event Text Entries

[138795877] (b)(4). The device involved has not been received for evaluation by the manufacturer at the time of this report. The device history record of batch number 74k1602223 that belongs to catalog number 41007 has been reviewed and no issues or discrepancies were found which could potentially be related to this complaint. Customer complaint cannot be confirmed based only on the information provided. To perform a proper and thorough investigation , it is necessary to evaluate the sample involved in this complaint. Root cause cannot be determined. Corrective actions cannot be established. If the sample becomes available this report will be updated with the evaluation results.
Patient Sequence No: 1, Text Type: N, H10


[138795878] Customer complaint reads: "for a patient under 15 liters, the balloon does not inflate or very little. Another mask was used but with the same issue. The patient desaturated until 77%. Using the other mask the saturation went back up to 88% but with difficulties. " (second event captured in mdr #3004365956-2019-00069). Clinical consequences reported "desaturation of the patient. A mask from an ambulance was used and the saturation went back up to normal and the balloon was perfectly inflated. I don't know what was the device used by ambulance. "
Patient Sequence No: 1, Text Type: D, B5


MAUDE Entry Details

Report Number3004365956-2019-00070
MDR Report Key8416642
Report SourceCOMPANY REPRESENTATIVE,FOREIG
Date Received2019-03-13
Date of Report2019-03-06
Date of Event2019-02-22
Date Mfgr Received2019-04-30
Device Manufacturer Date2016-10-19
Date Added to Maude2019-03-13
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Reporter OccupationPHARMACIST
Health Professional3
Initial Report to FDA3
Report to FDA3
Event Location3
Manufacturer ContactKATHARINE TARPLEY
Manufacturer Street3015 CARRINGTON MILL BLVD
Manufacturer CityMORRISVILLE NC 27560
Manufacturer CountryUS
Manufacturer Postal27560
Manufacturer Phone9194334854
Manufacturer G1TELEFLEX MEDICAL
Manufacturer StreetPARQUE INDUSTRIAL FINSA
Manufacturer CityNUEVO LAREDO 88275
Manufacturer CountryMX
Manufacturer Postal Code88275
Single Use3
Previous Use Code3
Event Type3
Type of Report3

Device Details

Brand NameHUDSON MASK,HIGH CONC,ADULT,INTL
Generic NameMASK, OXYGEN
Product CodeBYG
Date Received2019-03-13
Catalog Number41007
Lot Number74K1602223
Device AvailabilityY
Device Eval'ed by MfgrR
Device Sequence No1
Device Event Key0
ManufacturerTELEFLEX MEDICAL