LARGE METAPHYSIS RASP STRAIGHT N/A 00789201380

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed from a distributor,health profession report with the FDA on 2019-03-13 for LARGE METAPHYSIS RASP STRAIGHT N/A 00789201380 manufactured by Zimmer Biomet, Inc..

Event Text Entries

[138700178] (b)(4). Concomitant medical devices: versys fiber metal midcoat stem # item 00784101330 lot 64161727. Multiple mdr reports were filed for this event, please see associated reports: 0001822565-2019-01116. If any further information is found which would change or alter any conclusions or information, a supplemental will be filed accordingly. Zimmer biomet will continue to monitor for trends.
Patient Sequence No: 1, Text Type: N, H10


[138700179] It was reported that during a hip surgery, the broach fitted in the femur easily, but during implanting stem it was tight and the femur got cracked. Cables were applied to get the stem in. The rasp used and the actual implant had different profiles and the surgeon is suspecting that the rasp id of wrong size. A delay of 30-45 min occurred during the procedure for application of cable to reduce fracture. Attempts have been made and additional information on the reported event is unavailable at this time.
Patient Sequence No: 1, Text Type: D, B5


MAUDE Entry Details

Report Number0001822565-2019-01117
MDR Report Key8417678
Report SourceDISTRIBUTOR,HEALTH PROFESSION
Date Received2019-03-13
Date of Report2019-08-22
Date of Event2019-02-12
Date Mfgr Received2019-08-20
Device Manufacturer Date2001-02-01
Date Added to Maude2019-03-13
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Health Professional3
Initial Report to FDA3
Report to FDA3
Event Location3
Manufacturer ContactMS. CHRISTINA ARNT
Manufacturer Street56 E. BELL DR.
Manufacturer CityWARSAW IN 46582
Manufacturer CountryUS
Manufacturer Postal46582
Manufacturer Phone5745273773
Manufacturer G1ZIMMER BIOMET, INC.
Manufacturer Street56 E. BELL DRIVE
Manufacturer CityWARSAW IN 46582
Manufacturer CountryUS
Manufacturer Postal Code46582
Single Use3
Previous Use Code3
Event Type3
Type of Report3

Device Details

Brand NameLARGE METAPHYSIS RASP STRAIGHT
Generic NamePROSTHESIS HIP
Product CodeHTR
Date Received2019-03-13
Model NumberN/A
Catalog Number00789201380
Lot NumberNI
OperatorHEALTH PROFESSIONAL
Device AvailabilityN
Device Eval'ed by MfgrR
Device Sequence No1
Device Event Key0
ManufacturerZIMMER BIOMET, INC.
Manufacturer Address56 E. BELL DRIVE WARSAW IN 46582 US 46582


Patients

Patient NumberTreatmentOutcomeDate
101. Required No Informationntervention 2019-03-13