ADULT DUAL HEATED WIRE BREATHING SYSTEM 22MM 1.6M KIT 3 038-31-203U 038-32-203U

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed from a company representative report with the FDA on 2019-03-14 for ADULT DUAL HEATED WIRE BREATHING SYSTEM 22MM 1.6M KIT 3 038-31-203U 038-32-203U manufactured by Flexicare Medical Dongguan Ltd..

Event Text Entries

[138761339] Root cause identified as failure of the seal on the chamber float to function properly as a result of deformation caused by the welding process used during manufacturing. Action taken: training of the operators to identify correct/incorrect assembly/function of stem and seal holder. Introduction of go/no go gauges. New jig that reduces risk of deformation during assembly. 100% inspection on production line.
Patient Sequence No: 1, Text Type: N, H10


[138761340] The chamber in the 038-31-203u would not fill with water after spiking the bag. The water would not flow into the chamber even after trying to squeeze the water bag. The clip near the spike on the end of the tubing was unclipped.
Patient Sequence No: 1, Text Type: D, B5


MAUDE Entry Details

Report Number3006061749-2019-00007
MDR Report Key8419647
Report SourceCOMPANY REPRESENTATIVE
Date Received2019-03-14
Date of Report2019-03-14
Date of Event2019-02-14
Date Mfgr Received2019-02-15
Device Manufacturer Date2017-05-31
Date Added to Maude2019-03-14
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Reporter OccupationOTHER HEALTH CARE PROFESSIONAL
Health Professional3
Initial Report to FDA3
Report to FDA3
Event Location3
Manufacturer ContactMR RICHARD DOWNS
Manufacturer StreetCYNON VALLEY BUSINESS PARK
Manufacturer CityMOUNTAIN ASH, RCT CF454ER
Manufacturer CountryUK
Manufacturer PostalCF45 4ER
Manufacturer G1FLEXICARE MEDICAL DONGGUAN LTD.
Manufacturer StreetNO. B-15 XICHENG IND ZONE 1 HENGLI TOWN
Manufacturer CityDONGGUAN CITY, GUANGDONG 523460
Manufacturer CountryCH
Manufacturer Postal Code523460
Single Use3
Remedial ActionMA
Previous Use Code3
Event Type3
Type of Report3

Device Details

Brand NameADULT DUAL HEATED WIRE BREATHING SYSTEM 22MM 1.6M KIT 3
Generic NameHEATED WIRE VENTILATOR BREATHING SYSTEM
Product CodeBZO
Date Received2019-03-14
Model Number038-31-203U
Catalog Number038-32-203U
Lot Number170500532
OperatorHEALTH PROFESSIONAL
Device AvailabilityN
Device Eval'ed by MfgrR
Device Sequence No1
Device Event Key0
ManufacturerFLEXICARE MEDICAL DONGGUAN LTD.
Manufacturer AddressNO. B-15 XICHENG IND ZONE 1 HENGLI TOWN DONGGUAN CITY, GUANGDONG 523460 CH 523460


Patients

Patient NumberTreatmentOutcomeDate
101. Required No Informationntervention 2019-03-14

© 2026 FDA.report
This site is not affiliated with or endorsed by the FDA.