TDOC AIR-CHARGED CATHETER CAT 880

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed with the FDA on 2019-03-17 for TDOC AIR-CHARGED CATHETER CAT 880 manufactured by Laborie Medical Technologies Canada Ulc.

Event Text Entries

[139186684] This incident did not result in a serious injury of the patient or require medical intervention; however, this event is being reported to the fda based on a review of the risk documentation for the device which identified that were this malfunction to occur again, there is the possibility that medical intervention might be necessary.
Patient Sequence No: 1, Text Type: N, H10


[139186685] A laborie catheter which was used in a male patient in (b)(6) as part of a urodynamics study had a steel wire protruding from the catheter. The physician either did not notice the defect prior to inserting the catheter into the patient, or the defect occurred after insertion into the patient - we are unable to tell based on the information provided - which resulted in the patient's urethra receiving minor scratches upon removal of the catheter. No medical intervention was required as a result of this product malfunction; the patient was released and did not require additional follow-up or treatments.
Patient Sequence No: 1, Text Type: D, B5


MAUDE Entry Details

Report Number9681424-2019-00001
MDR Report Key8427288
Date Received2019-03-17
Date of Report2019-03-12
Date of Event2019-02-11
Date Mfgr Received2019-02-11
Date Added to Maude2019-03-17
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Health Professional3
Initial Report to FDA3
Report to FDA3
Event Location3
Manufacturer ContactMS. NICOLE BOSER
Manufacturer Street6415 NORTHWEST DRIVE UNIT 10
Manufacturer CityMISSISSAUGA, ONTARIO L4V1X1
Manufacturer CountryCA
Manufacturer PostalL4V1X1
Manufacturer G1CLINICAL INNOVATIONS LLC
Manufacturer Street747 W 4170 S
Manufacturer CityMURRAY UT 84123
Manufacturer CountryUS
Manufacturer Postal Code84123
Single Use3
Remedial ActionIN
Previous Use Code3
Event Type3
Type of Report0

Device Details

Brand NameTDOC AIR-CHARGED CATHETER
Generic NameCATHETER FOR USE WITH URODYNAMIC SYSTEM
Product CodeFEN
Date Received2019-03-17
Returned To Mfg2019-02-25
Model NumberCAT 880
Lot Number180716
OperatorHEALTH PROFESSIONAL
Device AvailabilityR
Device Eval'ed by MfgrY
Device Sequence No1
Device Event Key0
ManufacturerLABORIE MEDICAL TECHNOLOGIES CANADA ULC
Manufacturer Address6415 NORTHWEST DR UNIT 10 MISSISSAUGA, ONTARIO L4V1X1 CA L4V1X1


Patients

Patient NumberTreatmentOutcomeDate
10 2019-03-17

© 2026 FDA.report