CRYOMACS FREEZING BAGS 500 N/A 200-074-402

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed with the FDA on 2019-03-18 for CRYOMACS FREEZING BAGS 500 N/A 200-074-402 manufactured by Miltenyi Biotec Gmbh.

Event Text Entries

[139199266] The bag was stored directly in liquid nitrogen. So it's highly likely that liquid nitrogen entered the cryomacs bag through an insufficient welding seam and expanded after removal from this tank for thawing. If a bag is stored in the liquid phase of nitrogen it is recommended to place the bag into the vapor phase of nitrogen for a minimum of 4 hours before removal from the nitrogen tank and thawing. This is the second complaint regarding a breakage for this lot with a complaint rate of (b)(4).
Patient Sequence No: 1, Text Type: N, H10


[139199267] After removal from the liquid nitrogen tank and transport to the transplantation unit in liquid nitrogen, the overwrap bag expanded rapidly upon removal, which could be seen as an overwrap bubble protruding from the cassette- window. Upon opening of the cassette, the overwrap bag burst and the product was found to be ripped in two halves. The patient received a backup product for transplantation. For a potentially necessary second therapy no backup product would be available anymore.
Patient Sequence No: 1, Text Type: D, B5


MAUDE Entry Details

Report Number3005290010-2019-00001
MDR Report Key8427585
Date Received2019-03-18
Date of Report2019-03-14
Date of Event2019-02-15
Date Mfgr Received2019-02-18
Date Added to Maude2019-03-18
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Reporter OccupationOTHER HEALTH CARE PROFESSIONAL
Health Professional3
Initial Report to FDA3
Report to FDA3
Event Location3
Manufacturer ContactDR. BETTINA-JUDITH H
Manufacturer StreetFRIEDRICH-EBERT-STR. 68
Manufacturer CityBERGISCH GLADBACH, 51429
Manufacturer CountryGM
Manufacturer Postal51429
Manufacturer G1MILTENYI BIOTEC GMBH
Manufacturer StreetFRIEDRICH-EBERT-STR. 68
Manufacturer CityBERGISCH GLADBACH, 51429
Manufacturer CountryGM
Manufacturer Postal Code51429
Single Use3
Previous Use Code3
Event Type3
Type of Report0

Device Details

Brand NameCRYOMACS FREEZING BAGS 500
Generic NameCRYOMACS FREEZING BAGS 500
Product CodeKSR
Date Received2019-03-18
Model NumberN/A
Catalog Number200-074-402
Lot Number6180828034
OperatorHEALTH PROFESSIONAL
Device AvailabilityN
Device Eval'ed by MfgrR
Device Sequence No1
Device Event Key0
ManufacturerMILTENYI BIOTEC GMBH
Manufacturer AddressFRIEDRICH-EBERT-STR. 68 BERGISCH GLADBACH, 51429 GM 51429


Patients

Patient NumberTreatmentOutcomeDate
10 2019-03-18

© 2026 FDA.report