TOTAL PSA, TOTAL (FREE + COMPLEXED) PSA - PROSTATE-SPECIFIC ANTIGEN (TPSA) 04641655190

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed from a foreign,health professional,u report with the FDA on 2019-03-18 for TOTAL PSA, TOTAL (FREE + COMPLEXED) PSA - PROSTATE-SPECIFIC ANTIGEN (TPSA) 04641655190 manufactured by Roche Diagnostics.

Event Text Entries

[139087456] This event occurred in (b)(6).
Patient Sequence No: 1, Text Type: N, H10


[139087497] The initial reporter questioned low results for multiple patients tested for elecsys total psa immunoassay (total psa) on a cobas e 411 immunoassay analyzer. The customer provided an example for 1 patient with an initial result from the e411 analyzer of 0. 002 ng/ml with a data flag. The sample was repeated twice at a different laboratory using the cobas 6000 e 601 module with results of 0. 158 ng/ml and 0. 157 ng/ml. The repeat results were believed to be correct based on the previous results for this patient and that the patient had a recent prostate carcinoma operation. The results were reported outside of the laboratory. There was no allegation that an adverse event occurred. The e411 analyzer serial number was (b)(4). The customer pulled patient samples from the last 8 weeks (50 samples) and repeated them. Discrepant results were identified for all 50 samples. Refer to attached data for these results. The customer believes the issue was caused by a calibration issue from (b)(6) 2019. A field service engineer (fse) visited the customer site and performed new calibration and the customer has had no further issues.
Patient Sequence No: 1, Text Type: D, B5


MAUDE Entry Details

Report Number1823260-2019-01109
MDR Report Key8428346
Report SourceFOREIGN,HEALTH PROFESSIONAL,U
Date Received2019-03-18
Date of Report2019-04-29
Date of Event2019-02-01
Date Mfgr Received2019-02-27
Date Added to Maude2019-03-18
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Reporter OccupationOTHER HEALTH CARE PROFESSIONAL
Health Professional3
Initial Report to FDA3
Report to FDA3
Event Location3
Manufacturer ContactNA MICHAEL LESLIE
Manufacturer Street9115 HAGUE ROAD NA
Manufacturer CityINDIANAPOLIS IN 46250
Manufacturer CountryUS
Manufacturer Postal46250
Manufacturer Phone3175214343
Manufacturer G1ROCHE DIAGNOSTICS GMBH
Manufacturer StreetSANDHOFERSTRASSE 116 NA
Manufacturer CityMANNHEIM (BADEN-WURTTEMBERG) 68305
Manufacturer CountryGM
Manufacturer Postal Code68305
Single Use3
Previous Use Code3
Removal Correction NumberNA
Event Type3
Type of Report3

Device Details

Brand NameTOTAL PSA, TOTAL (FREE + COMPLEXED) PSA - PROSTATE-SPECIFIC ANTIGEN (TPSA)
Generic NameTOTAL,PROSTATE SPECIFIC ANTIGEN FOR DETECTION OF PROSTATE CANCER
Product CodeMTF
Date Received2019-03-18
Model NumberNA
Catalog Number04641655190
Lot Number35107401
OperatorHEALTH PROFESSIONAL
Device AvailabilityY
Device Eval'ed by MfgrR
Device Sequence No1
Device Event Key0
ManufacturerROCHE DIAGNOSTICS
Manufacturer Address9115 HAGUE ROAD NA INDIANAPOLIS IN 462500457 US 462500457


Patients

Patient NumberTreatmentOutcomeDate
10 2019-03-18