ANTI-TSHR, ANTIBODIES TO TSH RECEPTOR 07026951190

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed from a foreign,health professional,u report with the FDA on 2019-03-18 for ANTI-TSHR, ANTIBODIES TO TSH RECEPTOR 07026951190 manufactured by Roche Diagnostics.

Event Text Entries

[139102520] This event occurred in (b)(6).
Patient Sequence No: 1, Text Type: N, H10


[139102521] The customer complained of questionable elecsys tsh assay results tested on a cobas 8000 e 602 module and elecsys ft4 iii assay, elecsys ft3 iii, and elecsys anti-tshr immunoassay results tested on a cobas 8000 e 801 module for 1 patient sample compared to competitor methods. The e 801 serial number was (b)(4). The e 602 serial number was not provided. This medwatch will cover anti-tshr. Refer to medwatch with patient identifier (b)(6) for information on the ft4 iii results, medwatch with patient identifier (b)(6) for information on the ft3 iii results, and medwatch with patient identifier (b)(6) for information on the tsh results. The tsh result was 0. 18 iu/l on the e 602 and 1. 62 iu/l on the beckman dxi access. The ft4 iii result was >100 pmol/l on the e 801 and 16. 73 pmol/l on the siemens centaur. The ft3 iii result was 17. 1 pmol/l on the e 801 and 4. 915 pmol/l on the siemens centaur. The anti-tshr result was >40 iu/l on the e 801 and <0. 9 iu/l on the thermo fisher brahms kryptor. The results from the e 801 and the e 602 were reported outside of the laboratory and the physician questioned the results. There was no adverse event.
Patient Sequence No: 1, Text Type: D, B5


MAUDE Entry Details

Report Number1823260-2019-01123
MDR Report Key8429153
Report SourceFOREIGN,HEALTH PROFESSIONAL,U
Date Received2019-03-18
Date of Report2019-04-25
Date of Event2019-01-28
Date Mfgr Received2019-03-01
Date Added to Maude2019-03-18
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Reporter OccupationOTHER HEALTH CARE PROFESSIONAL
Health Professional3
Initial Report to FDA3
Report to FDA3
Event Location3
Manufacturer ContactNA MICHAEL LESLIE
Manufacturer Street9115 HAGUE ROAD NA
Manufacturer CityINDIANAPOLIS IN 46250
Manufacturer CountryUS
Manufacturer Postal46250
Manufacturer Phone3175214343
Manufacturer G1ROCHE DIAGNOSTICS GMBH
Manufacturer StreetSANDHOFERSTRASSE 116 NA
Manufacturer CityMANNHEIM (BADEN-WURTTEMBERG) 68305
Manufacturer CountryGM
Manufacturer Postal Code68305
Single Use3
Previous Use Code3
Removal Correction NumberNA
Event Type3
Type of Report3

Device Details

Brand NameANTI-TSHR, ANTIBODIES TO TSH RECEPTOR
Generic NameANTI-TSHR IMMUNOASSAY
Product CodeJZO
Date Received2019-03-18
Model NumberNA
Catalog Number07026951190
Lot Number35935501
OperatorHEALTH PROFESSIONAL
Device AvailabilityY
Device Eval'ed by MfgrR
Device Sequence No1
Device Event Key0
ManufacturerROCHE DIAGNOSTICS
Manufacturer Address9115 HAGUE ROAD NA INDIANAPOLIS IN 462500457 US 462500457


Patients

Patient NumberTreatmentOutcomeDate
10 2019-03-18