BIOGLUE SURGICAL ADHESIVE BG3502-5-G

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed from a foreign,user facility report with the FDA on 2019-03-18 for BIOGLUE SURGICAL ADHESIVE BG3502-5-G manufactured by Cryolife, Inc..

Event Text Entries

[139110273] This investigation is currently ongoing. Any additional information will be provided in the follow-up report.
Patient Sequence No: 1, Text Type: N, H10


[139110276] According to the initial report, [the surgeon] has done redo the patient, who had neuro surgery of spinal tumor at thoracic spine using bg 2ml for dural repair in last (b)(6), due to mass formation of bioglue. Additional information from the operating surgeon relays that on (b)(6) 2018 the patient had a meningioma (intradural extramedullary type) removed from the thoracic spine. The bioglue was applied after the dura was sutured with prolene. The area was dry where the bioglue was applied. The mass was located just over the dura mater. An mri (magnetic resonance imaging) was performed before surgery on (b)(6) 2019 to remove the bioglue mass. The bioglue mass was diagnosed about 3 months after the initial surgery. After the removal of the tumor, the patient felt leg weakness was improving. But the patient feels the leg weakness is becoming worse recently.
Patient Sequence No: 1, Text Type: D, B5


MAUDE Entry Details

Report Number1063481-2019-00010
MDR Report Key8429746
Report SourceFOREIGN,USER FACILITY
Date Received2019-03-18
Date of Report2019-06-07
Date of Event2018-12-31
Date Mfgr Received2019-03-09
Date Added to Maude2019-03-18
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Health Professional3
Initial Report to FDA3
Report to FDA3
Event Location3
Manufacturer ContactROCHELLE MANEY
Manufacturer Street1655 ROBERTS BLVD. NW
Manufacturer CityKENNESAW GA 30144
Manufacturer CountryUS
Manufacturer Postal30144
Manufacturer Phone7704193355
Manufacturer G1CRYOLIFE, INC.
Manufacturer Street1655 ROBERTS BLVD NW
Manufacturer CityKENNESAW GA 30144
Manufacturer CountryUS
Manufacturer Postal Code30144
Single Use3
Previous Use Code3
Event Type3
Type of Report3

Device Details

Brand NameBIOGLUE SURGICAL ADHESIVE
Generic NameGLUE, SURGICAL, ARTERIES
Product CodeMUQ
Date Received2019-03-18
Model NumberBG3502-5-G
Lot NumberUNKNOWN
Device AvailabilityN
Device Eval'ed by Mfgr*
Device Sequence No1
Device Event Key0
ManufacturerCRYOLIFE, INC.
Manufacturer Address1655 ROBERTS BLVD. NW KENNESAW GA 30144 US 30144


Patients

Patient NumberTreatmentOutcomeDate
101. Hospitalization; 2. Other; 3. Required No Informationntervention 2019-03-18