BREWER ACCESS EXAM TABLE 5000-04

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed with the FDA on 2019-03-18 for BREWER ACCESS EXAM TABLE 5000-04 manufactured by 2182356.

Event Text Entries

[139163390] According to the reporting person, a "rather large" patient placed his left hand on and somewhat under the upholstered top. When he sat on the table, his left hand middle finger was cut by something sharp, and his finger/nail fell on floor. The plastic surgical repair was the next morning using a skin graft of pig's skin. When i asked her to clarify "rather large", she stated (b)(6) lb. She stated that "apparently the bumpers fall off easily... After the incident, when we looked at the 37 tables in the clinic, we found 18 tables that were missing one or both of the bumpers. " she had their service company come in to replace the missing bumpers. The table was most recently inspected in january by their service company. Picture sent by the reporting person showed no bumper or screw on one side. The reporting person stated that there was not a specific sharp edge on the table. (refer to mw5081405).
Patient Sequence No: 1, Text Type: D, B5


MAUDE Entry Details

Report Number2182536-2018-00001
MDR Report Key8430505
Date Received2019-03-18
Date of Report2018-12-03
Date of Event2018-07-30
Date Mfgr Received2018-08-01
Device Manufacturer Date2002-07-17
Date Added to Maude2019-03-18
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Health Professional3
Initial Report to FDA3
Report to FDA3
Event Location3
Manufacturer ContactEMILY VANDE HEI
Manufacturer StreetN88 W13901 MAIN STREET SUITE 100
Manufacturer CityMENOMONEE FALLS WI 53051
Manufacturer CountryUS
Manufacturer Postal53051
Manufacturer Phone2622937164
Manufacturer G1THE BREWER COMPANY LLC
Manufacturer StreetN88 W13901 MAIN STREET SUITE 100
Manufacturer CityMENOMONEE FALLS WI 53051
Manufacturer CountryUS
Manufacturer Postal Code53051
Single Use3
Remedial ActionRP
Previous Use Code3
Event Type3
Type of Report0

Device Details

Brand NameBREWER ACCESS EXAM TABLE
Generic NameCHAIR, EXAMINATION AND TREATMENT
Product CodeFRK
Date Received2019-03-18
Model Number5000-04
Catalog Number5000-04
Lot NumberN/A
OperatorHEALTH PROFESSIONAL
Device AvailabilityN
Device Age16 YR
Device Eval'ed by MfgrR
Device Sequence No1
Device Event Key0
Manufacturer2182356
Manufacturer AddressN88 W13901 MAIN STREET SUITE 100 MENOMONEE FALLS WI 53051 US 53051


Patients

Patient NumberTreatmentOutcomeDate
101. Deathisabilit 2019-03-18