DT UNIVERSAL DRILL V040935000

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed with the FDA on 2019-03-19 for DT UNIVERSAL DRILL V040935000 manufactured by Recherches Techniques Dentaires (r.t.d.).

Event Text Entries

[140151440] There has been a previous report received with a similar device where this malfunction resulted in a serious injury. Therefore, it must be presumed that recurrence of this malfunction could possibly cause or contribute to a serious injury or require medical or surgical intervention to preclude such. As such, this event is reportable per 21cfr part 803. Investigation results received: the sem investigation did not show or detect defect in the material. According to vdw, the customer followed the procedure by using a canal root preparation and low speed as recommended 1000-200 pm. However in absence of clinical procedure details asked by e-mail dated on (b)(6) 2016, it is difficult to determine the exact reason. The dark/back at the broken part may suggest to think that the drill was blocked (for undetermined reason) when during the drilling operation. This submission is being made after an acquisition and internal audit of rtd.
Patient Sequence No: 1, Text Type: N, H10


[140151441] It was reported that a dt universal drill broke during use. The broken part was retrieved and no injury resulted.
Patient Sequence No: 1, Text Type: D, B5


MAUDE Entry Details

Report Number3001226627-2016-00001
MDR Report Key8434542
Date Received2019-03-19
Date of Report2019-03-19
Date Mfgr Received2016-02-25
Date Added to Maude2019-03-19
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Reporter OccupationDENTIST
Health Professional3
Initial Report to FDA3
Report to FDA3
Event Location3
Manufacturer ContactMR. KARL NITTINGER
Manufacturer Street221 W. PHILADELPHIA ST. SUITE 60W
Manufacturer CityYORK PA 17401
Manufacturer CountryUS
Manufacturer Postal17401
Manufacturer Phone7178494424
Manufacturer G1RECHERCHES TECHNIQUES DENTAIRES (R.T.D.)
Manufacturer Street3 RUE LOUIS NEEL
Manufacturer CityST. EGREVE ISERE, 38120
Manufacturer CountryFR
Manufacturer Postal Code38120
Single Use3
Previous Use Code3
Event Type3
Type of Report0

Device Details

Brand NameDT UNIVERSAL DRILL
Generic NameDRILL, DENTAL, INTRAORAL
Product CodeDZA
Date Received2019-03-19
Returned To Mfg2016-02-25
Model NumberNA
Catalog NumberV040935000
Lot NumberUNK
OperatorHEALTH PROFESSIONAL
Device AvailabilityR
Device Eval'ed by MfgrY
Device Sequence No1
Device Event Key0
ManufacturerRECHERCHES TECHNIQUES DENTAIRES (R.T.D.)
Manufacturer Address3 RUE LOUIS NEEL ST. EGREVE ISERE, 38120 FR 38120


Patients

Patient NumberTreatmentOutcomeDate
10 2019-03-19

© 2026 FDA.report
This site is not affiliated with or endorsed by the FDA.