DT LIGHT POST V861961000101

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed from a foreign,health professional report with the FDA on 2019-03-19 for DT LIGHT POST V861961000101 manufactured by Recherches Techniques Dentaires (r.t.d.).

Event Text Entries

[139321581] While no serious injury was reported, there has been a previous report received where breakage resulted in a serious injury. Therefore, it must be presumed that recurrence of this malfunction could possibly cause or contribute to a serious injury or require medical or surgical intervention to preclude such. As such, this event is reportable per 21cfr part 803. The broken post may probably due to an excessive force when cutting or using dull instrument. It is recommended to cust fibre posts with a diamond coated disc or a diamond bur. This submission is being made after an acquisition and internal audit of rtd.
Patient Sequence No: 1, Text Type: N, H10


[139321582] It was reported that a dt light post broke during the cut of the post, before inserting it in patient mouth.
Patient Sequence No: 1, Text Type: D, B5


MAUDE Entry Details

Report Number3001226627-2016-00002
MDR Report Key8434544
Report SourceFOREIGN,HEALTH PROFESSIONAL
Date Received2019-03-19
Date of Report2019-03-19
Date Mfgr Received2016-04-07
Date Added to Maude2019-03-19
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Reporter OccupationDENTIST
Health Professional3
Initial Report to FDA3
Report to FDA3
Event Location3
Manufacturer ContactMR. KARL NITTINGER
Manufacturer Street221 W. PHILADELPHIA ST. SUITE 60W
Manufacturer CityYORK PA 17401
Manufacturer CountryUS
Manufacturer Postal17401
Manufacturer Phone7178494424
Manufacturer G1RECHERCHES TECHNIQUES DENTAIRES (R.T.D.)
Manufacturer Street3 RUE LOUIS NEEL
Manufacturer CityST. EGREVE ISERE, 38120
Manufacturer CountryFR
Manufacturer Postal Code38120
Single Use3
Previous Use Code3
Event Type3
Type of Report3

Device Details

Brand NameDT LIGHT POST
Generic NamePOST, ROOT CANAL
Product CodeELR
Date Received2019-03-19
Returned To Mfg2016-05-09
Model NumberNA
Catalog NumberV861961000101
Lot Number61412
OperatorHEALTH PROFESSIONAL
Device AvailabilityR
Device Eval'ed by MfgrY
Device Sequence No1
Device Event Key0
ManufacturerRECHERCHES TECHNIQUES DENTAIRES (R.T.D.)
Manufacturer Address3 RUE LOUIS NEEL ST. EGREVE ISERE, 38120 FR 38120


Patients

Patient NumberTreatmentOutcomeDate
10 2019-03-19

© 2024 FDA.report